CRCST Practice Test: 50 Free Questions With Answers
This free CRCST practice test has 50 original, unofficial questions covering all seven domains of HSPA's current CRCST content outline. It is not a full-length exam. Pick an answer, then open the explanation: each one names a source you can check, and there's no signup.
Practice questions
Question 1 of 50
Cleaning, Decontamination and Disinfection · CRCST-001
A reusable laryngoscope blade touches the patient's mucous membranes but does not enter sterile tissue. Under the Spaulding classification, what is the minimum level of processing it needs between patients?
- A. Intermediate-level disinfection
- B. Cleaning with detergent only
- C. High-level disinfection
- D. Low-level disinfection
Show answer and explanation
Correct answer: C. Items that contact mucous membranes or nonintact skin are semicritical. CDC lists laryngoscope blades as semicritical items and says they minimally require high-level disinfection.
Why the other choices don't fit
- A: Intermediate-level disinfection is not the minimum CDC sets for semicritical items.
- B: Cleaning removes soil but does not disinfect. It is the step before disinfection, not a substitute for it.
- D: Low-level disinfection is for noncritical items that touch only intact skin.
Source: CDC, Guideline for Disinfection and Sterilization (2008), "A Rational Approach to Disinfection and Sterilization", "Semicritical Items"; CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 3.b
Question 2 of 50
Sterilization Process · CRCST-002
In a gravity-displacement steam sterilizer at 250°F (121°C), what is the minimum exposure time CDC lists for wrapped instruments?
- A. 4 minutes
- B. 10 minutes
- C. 15 minutes
- D. 30 minutes
Show answer and explanation
Correct answer: D. CDC's Table 7 lists 30 minutes at 121°C for wrapped instruments in a gravity-displacement sterilizer.
Why the other choices don't fit
- A: 4 minutes is the dynamic-air-removal (prevacuum) time at 270°F (132°C).
- B: 10 minutes is not a Table 7 minimum for this cycle.
- C: 15 minutes is the gravity time for wrapped instruments at the higher 270°F (132°C) setting.
Source: CDC, Table 7: Minimum cycle times for steam sterilization cycles, rows for gravity displacement
Question 3 of 50
Departmental Considerations · CRCST-003
Which air pressure relationship should the decontamination area have compared with the clean areas next to it?
- A. Positive pressure, so clean air pushes out of decontamination
- B. Neutral pressure, so no air moves between rooms
- C. Negative pressure, so air flows into decontamination and contaminants stay there
- D. Whatever pressure the sterile storage room has
Show answer and explanation
Correct answer: C. CDC says the decontamination area should contain contaminants and limit their flow to clean areas. The American Institute of Architects guidance CDC cites calls for negative pressure in decontamination.
Why the other choices don't fit
- A: Positive pressure would push contaminated air out into clean areas.
- B: Neutral pressure gives no directional protection.
- D: Sterile storage has the opposite goal: keeping contaminants out, not in.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Physical Facilities"
Question 4 of 50
Preparation and Packaging · CRCST-004
You are assembling a tray of ring-handled instruments for steam sterilization. How should the hinged instruments be positioned?
- A. Opened only if they are made of stainless steel
- B. Closed and locked on the first ratchet so they take less space
- C. Opened (unlocked), with removable parts disassembled unless the manufacturer says otherwise
- D. Closed and stacked tightly so they don't shift
Show answer and explanation
Correct answer: C. CDC, summarizing AAMI and professional guidance, says hinged instruments should be opened and items with removable parts disassembled unless the device manufacturer provides instructions or test data to the contrary. Steam has to reach the box locks and inner surfaces.
Why the other choices don't fit
- A: The rule is about sterilant contact, not the metal type.
- B: A locked box lock hides surfaces from the sterilant.
- D: Tight stacking blocks steam contact and can damage instruments.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Packaging"
Question 5 of 50
Sterilization Process · CRCST-005
When must a Bowie-Dick (air-removal) test be run on a prevacuum steam sterilizer?
- A. Only after major repairs
- B. Once a week, in a full load
- C. Only when the printout shows a problem
- D. Each day the sterilizer is used, in an empty chamber before the first processed load
Show answer and explanation
Correct answer: D. CDC says the Bowie-Dick test is run each day a vacuum-type steam sterilizer is used, before the first processed load, with the test pack in an otherwise empty chamber.
Why the other choices don't fit
- A: Testing after repair is required too, but it does not replace the daily test.
- B: Weekly testing in a full load describes routine biological indicator monitoring, not the daily air-removal test.
- C: The test is a scheduled daily check, not a response to a printout problem.
Source: CDC, Guideline for Disinfection and Sterilization (2008), "Steam Sterilization", "Overview" (Bowie-Dick test); CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Monitoring"
Question 6 of 50
Cleaning, Decontamination and Disinfection · CRCST-006
Which statement about enzymatic cleaners is correct?
- A. They help break down protein soil such as blood but are not disinfectants
- B. They should be reused all shift to save chemicals
- C. They work best when instruments are left to dry first
- D. They disinfect instruments, so a separate disinfection step can be skipped
Show answer and explanation
Correct answer: A. Enzymes, usually proteases, attack proteins in soil like blood and pus. CDC states enzymatic cleaners are not disinfectants and recommends discarding them after each use because they do not slow microbial growth.
Why the other choices don't fit
- B: CDC recommends discarding enzymatic cleaners after each use.
- C: Dried soil is harder to remove. CDC says to clean as soon as practical after use.
- D: Enzymatic cleaners have no disinfectant claim.
Source: CDC, Guideline for Disinfection and Sterilization (2008), "Cleaning", "Cleaning"; CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Recs. 2.b.ii and 7.f
Question 7 of 50
Sterile Storage, Transport and Inventory Management · CRCST-007
Sterile supplies are being placed on open shelving. Which spacing matches CDC's storage guidance?
- A. At least 2 inches from the floor and touching the ceiling
- B. 8 to 10 inches from the floor, 5 inches from the ceiling (18 inches from a sprinkler head), and 2 inches from outside walls
- C. 12 inches from the floor and 12 inches from all walls
- D. On the floor if the packages are in plastic
Show answer and explanation
Correct answer: B. CDC says sterile supplies should be stored 8 to 10 inches from the floor, 5 inches from the ceiling (18 inches if near a sprinkler head) and 2 inches from outside walls, for air circulation, cleaning and fire-code compliance.
Why the other choices don't fit
- A: Two inches from the floor is too low, and nothing should touch the ceiling.
- C: These are not the distances CDC gives.
- D: Plastic packaging does not make floor storage acceptable.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Storage"
Question 8 of 50
Sterilization Process · CRCST-008
Which microorganism's spores are used in biological indicators for steam sterilization, and at what temperature are they incubated?
- A. Staphylococcus aureus, incubated at 37°C
- B. Geobacillus stearothermophilus, incubated at 55–60°C
- C. Clostridioides difficile, incubated at 25°C
- D. Bacillus atrophaeus, incubated at 35–37°C
Show answer and explanation
Correct answer: B. CDC states G. stearothermophilus spores are used to monitor steam sterilization (and hydrogen peroxide gas plasma and liquid peracetic acid) and are incubated at 55–60°C.
Why the other choices don't fit
- A: S. aureus is not a spore-former and is not used in sterilizer biological indicators.
- C: C. difficile is not a sterilizer test organism.
- D: B. atrophaeus monitors ethylene oxide and dry heat and is incubated at 35–37°C.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Monitoring"
Question 9 of 50
Departmental Considerations · CRCST-009
A technician has weeping dermatitis on both hands. According to CDC, what should happen?
- A. Exclude the technician from direct contact with patient-care equipment
- B. Keep working in decontamination with double gloves
- C. Allow work in any area as long as hand lotion is used
- D. Assign the technician only to sterile storage without restriction
Show answer and explanation
Correct answer: A. CDC recommends excluding healthcare workers with weeping dermatitis of the hands from direct contact with patient-care equipment.
Why the other choices don't fit
- B: Gloves do not remove the risk from weeping skin lesions; CDC's recommendation is exclusion from direct contact.
- C: Lotion does not address the infection risk.
- D: Handling sterile packs is still direct contact with patient-care equipment.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 1.e
Question 10 of 50
Preparation and Packaging · CRCST-010
Steam-sterilized trays keep coming out of the sterilizer with moisture inside. Which tray factor does CDC name as a cause of wet packs?
- A. Heavy metal mass in the set
- B. Opening hinged instruments
- C. Using count sheets
- D. Placing a chemical indicator inside
Show answer and explanation
Correct answer: A. CDC notes that heavy metal mass is a cause of wet packs, meaning moisture inside the case and tray after the cycle. Wrap density and set design also affect drying.
Why the other choices don't fit
- B: Opening hinged instruments helps sterilant contact. It does not cause wet packs.
- C: Count sheets don't affect drying.
- D: An internal indicator is recommended and does not cause moisture.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Packaging"
Question 11 of 50
Cleaning, Decontamination and Disinfection · CRCST-011
How does an ultrasonic cleaner remove soil from instruments?
- A. By cavitation: sound waves create tiny bubbles that implode and loosen soil
- B. By spraying high-pressure jets through lumens
- C. By chemically sterilizing the surface
- D. By heating instruments to sterilizing temperatures
Show answer and explanation
Correct answer: A. CDC explains that ultrasonic cleaning removes soil by cavitation and implosion, as acoustic energy travels through the solution and disrupts the bonds that hold particles to surfaces.
Why the other choices don't fit
- B: Pressurized fluid through channels is fluidics, used in manual cleaning.
- C: CDC notes ultrasound alone does not significantly inactivate bacteria.
- D: Ultrasonic cleaners clean; they don't reach sterilizing temperatures.
Source: CDC, Guideline for Disinfection and Sterilization (2008), "Cleaning", "Cleaning"
Question 12 of 50
Departmental Considerations · CRCST-012
Where on a Safety Data Sheet (SDS) would you find the recommended personal protective equipment for a cleaning chemical?
- A. Section 4, First-Aid Measures
- B. Section 6, Accidental Release Measures
- C. Section 7, Handling and Storage
- D. Section 8, Exposure Controls/Personal Protection
Show answer and explanation
Correct answer: D. Under OSHA's Hazard Communication Standard, SDS Section 8 lists exposure limits, engineering controls and personal protective measures such as the types of eye, face, skin or respiratory protection.
Why the other choices don't fit
- A: Section 4 covers first aid after exposure.
- B: Section 6 covers spill and leak response.
- C: Section 7 covers safe handling and storage conditions.
Source: OSHA Brief, Hazard Communication Standard: Safety Data Sheets (OSHA 3514), Sections 4, 6, 7 and 8
Question 13 of 50
Sterilization Process · CRCST-013
A sterilizer's biological indicator test pack goes in a full load. Where in the chamber does CDC say it should be placed?
- A. Anywhere, as long as it lies flat
- B. Front, bottom section, near the drain
- C. Center of the top rack
- D. Top shelf, at the back
Show answer and explanation
Correct answer: B. CDC says to place the test pack flat in an otherwise fully loaded chamber in the area least favorable to sterilization, normally the front, bottom section near the drain.
Why the other choices don't fit
- A: Placement matters: the pack belongs in the most challenging location.
- C: Center-top is not the hardest spot for steam to reach.
- D: This is not the area CDC describes as the greatest challenge.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Monitoring"
Question 14 of 50
Preparation and Packaging · CRCST-014
Which wrapping technique creates a package within a package?
- A. Sequential wrapping
- B. Peel pouch with a single seal
- C. Nonsequential (simultaneous) wrapping
- D. Rigid container with a filter
Show answer and explanation
Correct answer: A. CDC describes sequential wrapping as using two sheets of sterilization wrap, one wrapped after the other, which creates a package within a package. Nonsequential wrapping uses two sheets at the same time.
Why the other choices don't fit
- B: A peel pouch is a different packaging system.
- C: Nonsequential wrapping folds two sheets together in one wrapping step.
- D: A rigid container is not a wrap.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Packaging"
Question 15 of 50
Cleaning, Decontamination and Disinfection · CRCST-015
You are emptying a tray of used instruments that includes reusable sharp hooks and blades. What is the safe way to retrieve them?
- A. Reach in carefully with heavy utility gloves
- B. Use forceps or another mechanical device instead of your hands
- C. Pour the tray out onto the counter
- D. Ask the OR to remove the sharps after the tray arrives
Show answer and explanation
Correct answer: B. OSHA prohibits processing contaminated reusable sharps in a way that requires employees to reach by hand into the containers holding them. CDC adds that employees should use engineering controls such as forceps.
Why the other choices don't fit
- A: Gloves don't make hand-reaching into sharps acceptable.
- C: Dumping instruments creates sharps and splash hazards.
- D: The tray has already arrived. You still have to retrieve the sharps safely.
Source: OSHA, Bloodborne Pathogens Standard, 29 CFR 1910.1030, 1910.1030(d)(4)(ii)(E); CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Cleaning"
Question 16 of 50
Sterilization Process · CRCST-016
A prevacuum steam sterilizer at 270°F (132°C) is processing wrapped instruments. What minimum exposure time does CDC's Table 7 list?
- A. 3 minutes
- B. 4 minutes
- C. 15 minutes
- D. 30 minutes
Show answer and explanation
Correct answer: B. CDC's Table 7 lists a 4-minute exposure at 132°C for wrapped instruments in a dynamic-air-removal (prevacuum) sterilizer.
Why the other choices don't fit
- A: 3 minutes is not the Table 7 minimum for this cycle.
- C: 15 minutes is the gravity-displacement time at 132°C.
- D: 30 minutes is the gravity-displacement time at 121°C.
Source: CDC, Table 7: Minimum cycle times for steam sterilization cycles, row for dynamic-air-removal, wrapped instruments
Question 17 of 50
Sterile Storage, Transport and Inventory Management · CRCST-017
Your department uses event-related shelf life. A wrapped tray sterilized four months ago is clean, dry and intact. What does that mean?
- A. It can be used only if a biological indicator is run again
- B. It can be used, because sterility is lost through events like tears, moisture or broken seals, not time alone
- C. It must be resterilized every 90 days
- D. It expired after 30 days and must be reprocessed
Show answer and explanation
Correct answer: B. Event-related shelf life treats a package as sterile until an event such as a tear, wetness or a broken seal compromises it. CDC says packaged sterile items can be used indefinitely under event-related storage unless the packaging is compromised.
Why the other choices don't fit
- A: Biological indicators monitor sterilizer cycles; they are not run on stored packages.
- C: No 90-day cycle applies under event-related shelf life.
- D: A fixed 30-day date is time-related practice, not event-related practice.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Storage"; CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 18.d
Question 18 of 50
Departmental Considerations · CRCST-018
A new decontamination technician will handle blood-contaminated instruments. By when must the employer make the hepatitis B vaccine available under OSHA's Bloodborne Pathogens Standard?
- A. Within 10 working days of initial assignment, after required training
- B. After one year of employment
- C. Only if the technician requests it in writing
- D. Only after a first exposure incident
Show answer and explanation
Correct answer: A. OSHA requires hepatitis B vaccination to be made available after required training and within 10 working days of initial assignment to employees with occupational exposure, unless an exception applies.
Why the other choices don't fit
- B: The standard sets 10 working days, not one year.
- C: The employer must offer it. The employee does not have to request it first.
- D: Waiting for an exposure defeats the purpose of preventive vaccination.
Source: OSHA, Bloodborne Pathogens Standard, 29 CFR 1910.1030, 1910.1030(f)(2)(i)
Question 19 of 50
Preparation and Packaging · CRCST-019
Which packaging material should NOT be used for items going into a hydrogen peroxide gas plasma sterilizer?
- A. Packaging cleared for that sterilizer
- B. Rigid containers validated for that cycle
- C. Linen and paper
- D. Any material the sterilizer manufacturer approves
Show answer and explanation
Correct answer: C. CDC lists linens and paper as unacceptable packaging for hydrogen peroxide gas plasma. Follow the sterilizer and packaging manufacturers' compatibility instructions.
Why the other choices don't fit
- A: Cleared, compatible packaging is what you should use.
- B: A container validated for the cycle is acceptable.
- D: Manufacturer-approved materials are the correct choice.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Packaging"
Question 20 of 50
Cleaning, Decontamination and Disinfection · CRCST-020
When a washer-disinfector is loaded, which practice does CDC recommend?
- A. Close box locks to protect them from spray
- B. Open hinged instruments fully and disassemble instruments as much as possible
- C. Stack instruments to fit more in each cycle
- D. Keep instruments in closed peel pouches
Show answer and explanation
Correct answer: B. CDC says hinged instruments should be opened fully for adequate contact with the detergent solution, stacking should be avoided and instruments should be disassembled as much as possible.
Why the other choices don't fit
- A: Closed box locks trap soil.
- C: Stacking blocks spray and detergent contact.
- D: Closed packaging prevents cleaning.
Source: CDC, Guideline for Disinfection and Sterilization (2008), "Cleaning", "Cleaning"
Question 21 of 50
Sterilization Process · CRCST-021
A steam sterilizer has just had a major repair. Before it goes back into routine use, what testing does CDC describe?
- A. One loaded cycle with a chemical indicator
- B. A single Bowie-Dick test
- C. No testing if the repair technician signs off
- D. Three consecutive empty cycles with biological and chemical indicators (plus Bowie-Dick tests for a prevacuum unit), all with acceptable results
Show answer and explanation
Correct answer: D. After installation, relocation, major repair or a sterilization failure, CDC describes three consecutive empty steam cycles with a biological and chemical indicator, and three Bowie-Dick tests in a prevacuum sterilizer. The unit stays out of use until all BIs are negative and CIs show the correct end point.
Why the other choices don't fit
- A: One cycle with only a chemical indicator doesn't qualify the sterilizer.
- B: A Bowie-Dick test checks air removal only and is not enough on its own.
- C: A repair sign-off doesn't replace verification testing.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Sterilization Cycle Verification"
Question 22 of 50
Sterile Storage, Transport and Inventory Management · CRCST-022
A sterile item in a heat-sealed, impervious plastic pouch falls to the floor. The seal is intact and the package isn't damaged. According to CDC, what should happen?
- A. Wipe it with alcohol and send it to the OR immediately
- B. Inspect it. If the package is undamaged and the seal intact, it can be considered not contaminated
- C. Return it to decontamination to be reprocessed
- D. Discard it
Show answer and explanation
Correct answer: B. CDC says a dropped package must be inspected for damage. If it is heat-sealed in impervious plastic and the seal is intact, it should be considered not contaminated, and undamaged items packaged in plastic need not be reprocessed.
Why the other choices don't fit
- A: Skipping inspection is the problem; alcohol wiping isn't the decision point.
- C: Reprocessing isn't required when the package passes inspection.
- D: An intact, undamaged plastic package doesn't need to be discarded.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Storage"
Question 23 of 50
Preparation and Packaging · CRCST-023
What does an external chemical indicator on a wrapped tray tell you?
- A. The steam penetrated to the center of the tray
- B. The package went through a sterilization process; it doesn't prove sterility
- C. The biological indicator passed
- D. The contents are sterile
Show answer and explanation
Correct answer: B. CDC says external chemical indicators show that a package has been processed through a sterilization cycle but do not prove sterilization. An internal indicator is preferred to verify sterilant penetration.
Why the other choices don't fit
- A: Penetration into the pack is what an internal indicator checks.
- C: Biological indicators are a separate test.
- D: No chemical indicator proves sterility.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Monitoring"
Question 24 of 50
Departmental Considerations · CRCST-024
Which practice does CDC say flash (immediate-use) steam sterilization should NOT be used for?
- A. Cleaned items placed in a container designed for that cycle
- B. Saving time or replacing the purchase of additional instrument sets
- C. Items that cannot be packaged, sterilized and stored before use
- D. Reprocessing an instrument dropped during a case that is needed right away
Show answer and explanation
Correct answer: B. CDC recommends against using flash sterilization for convenience, as an alternative to buying more instrument sets or to save time. It may be used when necessary for items needed immediately, such as a dropped instrument.
Why the other choices don't fit
- A: Using FDA-cleared containers designed for the cycle is part of doing it correctly.
- C: CDC lists items that can't be packaged, sterilized and stored before use as an acceptable use.
- D: CDC lists a dropped instrument that is needed immediately as an acceptable use.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Recs. 13.b, 13.d–13.f; CDC, Guideline for Disinfection and Sterilization (2008), "Flash Sterilization", "Uses"
Question 25 of 50
Cleaning, Decontamination and Disinfection · CRCST-025
A blood pressure cuff touches only a patient's intact skin. Which processing level does CDC recommend?
- A. Sterilization
- B. No processing ever needed
- C. High-level disinfection
- D. Low-level disinfection
Show answer and explanation
Correct answer: D. Noncritical equipment that touches intact skin, such as a blood pressure cuff, needs low-level disinfection.
Why the other choices don't fit
- A: Sterilization is required for critical items that enter sterile tissue or the vascular system.
- B: CDC still calls for low-level disinfection when the cuff is visibly soiled and on a regular schedule.
- C: High-level disinfection is the minimum for semicritical items.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Recs. 3.c and 4.c; CDC, Guideline for Disinfection and Sterilization (2008), "A Rational Approach to Disinfection and Sterilization", "Noncritical Items"
Question 26 of 50
Sterilization Process · CRCST-026
Which load requires a biological indicator in every load, with items quarantined until the result is negative when possible?
- A. Loads processed on weekends
- B. Loads of basins and utensils
- C. Loads of textile packs
- D. Any load containing implantable items
Show answer and explanation
Correct answer: D. CDC recommends using biological indicators for every load containing implantable items and quarantining those items, whenever possible, until the biological indicator is negative.
Why the other choices don't fit
- A: The day of the week doesn't change BI requirements.
- B: Basins don't trigger the every-load rule by themselves.
- C: Textile packs follow routine BI monitoring (at least weekly).
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 16.g; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Monitoring"
Question 27 of 50
Preparation and Packaging · CRCST-027
Which packaging materials does CDC say are unacceptable for ethylene oxide (EO) sterilization?
- A. Foil, polyvinyl chloride and kitchen-type plastic wrap
- B. Paper-plastic peel pouches cleared for EO
- C. Nonwoven wraps cleared for EO
- D. Rigid containers designed for EO
Show answer and explanation
Correct answer: A. CDC lists foil, polyvinyl chloride and polyvinylidene chloride (kitchen-type transparent wrap) as unacceptable packaging for EO.
Why the other choices don't fit
- B: Cleared peel pouches are acceptable choices.
- C: Cleared nonwoven wraps are acceptable choices.
- D: Containers designed and cleared for EO are acceptable.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Packaging"
Question 28 of 50
Cleaning, Decontamination and Disinfection · CRCST-028
A liquid high-level disinfectant is reused in a soaking basin. How often does CDC recommend testing its minimum effective concentration (MEC)?
- A. Only when the bottle is first opened
- B. Only when the solution looks cloudy
- C. Once a week
- D. Each day of use (or more often), with results documented
Show answer and explanation
Correct answer: D. CDC recommends testing the liquid sterilant or high-level disinfectant each day of use, or more often, with the appropriate chemical indicator, documenting results, and discarding it if it falls below the MEC or reaches the manufacturer's reuse-life limit.
Why the other choices don't fit
- A: Concentration drops with use, so one test at opening isn't enough.
- B: Appearance can't confirm concentration.
- C: Weekly testing is less frequent than CDC recommends.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 7.aa
Question 29 of 50
Sterilization Process · CRCST-029
A single steam-sterilizer biological indicator is positive. The mechanical printout and chemical indicators looked normal. What does CDC recommend first?
- A. Ignore it, because the chemical indicators passed
- B. Repeat the spore test immediately; non-implant items generally don't need recall unless the sterilizer or procedure is found defective
- C. Recall every item processed in the sterilizer in the past year
- D. Permanently remove the sterilizer from service
Show answer and explanation
Correct answer: B. CDC says a single positive spore test with steam probably doesn't indicate malfunction when mechanical and chemical monitors were acceptable, but the test should be repeated immediately. Objects other than implants need not be recalled unless the sterilizer or procedure is defective. If spore tests stay positive, stop using the sterilizer and recall the implicated loads.
Why the other choices don't fit
- A: A positive BI can't be ignored. It has to be retested.
- C: A one-year recall goes far beyond CDC's protocol.
- D: The sterilizer comes out of use if tests remain positive; one result alone doesn't mean permanent removal.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Monitoring"; CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 16.f
Question 30 of 50
Departmental Considerations · CRCST-030
Which statement about reprocessing a single-use device in a hospital is correct under FDA's approach as described by CDC?
- A. Only the original manufacturer can be inspected by FDA
- B. Reprocessed single-use devices are exempt from regulation
- C. The hospital is treated as the device manufacturer and held to the same standards
- D. Single-use devices can be reprocessed freely once they look clean
Show answer and explanation
Correct answer: C. CDC notes that FDA considers a hospital that reprocesses a single-use device to be the manufacturer of that device and regulates it by the same standards as the original manufacturer.
Why the other choices don't fit
- A: A reprocessing hospital is regulated as a manufacturer.
- B: Reprocessing doesn't remove regulatory requirements.
- D: Visual cleanliness doesn't satisfy FDA requirements.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 20.a
Question 31 of 50
Preparation and Packaging · CRCST-031
During inspection, a technician must decide whether a tray of instruments is ready to package. What is the minimum standard CDC states?
- A. Inspection happens only after sterilization
- B. A random sample of instruments is checked
- C. Instruments that went through the washer don't need inspection
- D. Every instrument is individually inspected and visibly clean
Show answer and explanation
Correct answer: D. CDC says that, at a minimum, all instruments should be individually inspected and be visibly clean. No real-time clinical test confirms cleaning, which is why inspection matters.
Why the other choices don't fit
- A: Inspection belongs before packaging. Soil can block sterilization.
- B: Sampling misses dirty instruments.
- C: Mechanical cleaning doesn't remove the need to inspect.
Source: CDC, Guideline for Disinfection and Sterilization (2008), "Cleaning", "Cleaning"
Question 32 of 50
Sterilization Process · CRCST-032
How long does CDC recommend aerating polyvinyl chloride tubing after ethylene oxide sterilization in a mechanical aerator?
- A. 30 minutes at room temperature
- B. 12 hours at 50°C or 8 hours at 60°C
- C. 2 hours at 37°C
- D. No aeration is needed
Show answer and explanation
Correct answer: B. CDC recommends complete aeration of EO-sterilized items and gives the example of PVC tubing needing 12 hours at 50°C or 8 hours at 60°C. Ambient room aeration takes about 7 days at 20°C.
Why the other choices don't fit
- A: Far too short. EO residues can harm patients.
- C: Not the CDC example.
- D: EO is absorbed by many materials, so aeration is required.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 14.e; CDC, Guideline for Disinfection and Sterilization (2008), "Ethylene Oxide 'Gas' Sterilization", "Overview"
Question 33 of 50
Cleaning, Decontamination and Disinfection · CRCST-033
Manual cleaning of a lumened instrument depends on two essential components. What are they?
- A. Friction and fluidics
- B. Disinfectant and drying time
- C. Heat and pressure
- D. Ultrasound and lubricant
Show answer and explanation
Correct answer: A. CDC names friction (scrubbing with a brush) and fluidics (fluid under pressure, used to clear internal channels) as the two essential components of manual cleaning.
Why the other choices don't fit
- B: Disinfection follows cleaning; it isn't how soil is removed.
- C: Heat and pressure describe steam sterilization.
- D: Ultrasound is mechanical cleaning, and lubricant comes after cleaning.
Source: CDC, Guideline for Disinfection and Sterilization (2008), "Cleaning", "Cleaning"
Question 34 of 50
Preparation and Packaging · CRCST-034
Which information does CDC recommend on the label of a sterilized item?
- A. The patient's diagnosis
- B. A load number showing the sterilizer used, the cycle or load number, the date of sterilization and, if applicable, the expiration date
- C. Only the word 'sterile'
- D. The surgeon's name only
Show answer and explanation
Correct answer: B. CDC recommends labeling sterilized items with a load number that identifies the sterilizer, the cycle or load number, the sterilization date and, if applicable, the expiration date. This supports recalls and traceability.
Why the other choices don't fit
- A: Patient diagnoses don't belong on instrument labels.
- C: A generic label can't trace the item to a load.
- D: The surgeon's name doesn't allow load tracing.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 18.c
Question 35 of 50
Sterile Storage, Transport and Inventory Management · CRCST-035
Which stock rotation method keeps older sterile supplies from sitting on the shelf too long?
- A. Last in, first out
- B. Use whatever is closest to the door
- C. Rotate stock once a year
- D. First in, first out (FIFO), using the oldest stock first
Show answer and explanation
Correct answer: D. Stock rotation works on first in, first out (FIFO): the oldest stock is distributed first so packs don't sit on the shelf longer than needed. Sterile-storage audit tools check that supplies are distributed on a FIFO basis.
Why the other choices don't fit
- A: Last in, first out leaves older stock sitting on the shelf.
- B: Convenience picking ignores dates.
- C: Annual rotation lets stock sit too long between checks.
Source: Asia Pacific Society of Infection Control (APSIC), CSSD Center of Excellence Audit Checklist, Section E: Sterile Storage and Distribution, Section E, Question 10; Namibia Ministry of Health and Social Services, CSSD Guideline (October 2023), published by WHO AFRO, §3.6, Commercially prepared items
Question 36 of 50
Sterilization Process · CRCST-036
Which steam sterilizer loading practice is correct?
- A. Pack items tightly to use the whole chamber
- B. Lay peel pouches flat and stack them
- C. Place peel pouches on edge in perforated or mesh-bottom baskets, and stand nonperforated basins on edge
- D. Place perforated trays at an angle to the shelf
Show answer and explanation
Correct answer: C. CDC lists loading principles: allow sterilant circulation, place perforated trays parallel to the shelf, place nonperforated containers such as basins on edge, load small items loosely in wire baskets and set peel packs on edge in perforated or mesh racks.
Why the other choices don't fit
- A: Tight packing blocks sterilant contact.
- B: Stacked flat pouches block steam circulation and drying.
- D: CDC says perforated trays go parallel to the shelf.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Loading"; CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 17.a
Question 37 of 50
Patient Care Equipment and Distribution · CRCST-037
While cleaning an infusion pump returned from a unit, you notice the power cord's outer jacket is cracked and the insulation is showing. What should you do?
- A. Wrap the cord in electrical tape and return the pump to service
- B. Use it only for short-term rentals
- C. Remove the pump from service and send it for repair, following your facility's process
- D. Return it to the unit and ask the nurse to watch the cord
Show answer and explanation
Correct answer: C. OSHA requires cord-and-plug equipment to be visually inspected for external defects such as damaged jackets or insulation, and a defective or damaged item must be removed from service until repaired.
Why the other choices don't fit
- A: Tape isn't a repair, and the defect may expose users to injury.
- B: A damaged cord is unsafe for any use.
- D: Sending damaged equipment back passes the hazard to the unit.
Source: OSHA, Electrical safety-related work practices, 29 CFR 1910.334, 1910.334(a)(2)(i)–(ii)
Question 38 of 50
Professional Development and Human Relation Skills · CRCST-038
A requisition reads 'cholecystectomy tray.' What does the suffix -ectomy mean?
- A. Making a cut into
- B. Viewing with an instrument
- C. Removal of
- D. Inflammation of
Show answer and explanation
Correct answer: C. MedlinePlus explains that -ectomy means removal of. A cholecystectomy is removal of the gallbladder.
Why the other choices don't fit
- A: Making a cut into is -otomy.
- B: Viewing with an instrument is -scopy.
- D: Inflammation is -itis.
Source: MedlinePlus (National Library of Medicine), Understanding Medical Words tutorial (PDF), word-endings section
Question 39 of 50
Cleaning, Decontamination and Disinfection · CRCST-039
Which statement about prions, the agents that cause Creutzfeldt-Jakob disease (CJD), is correct according to CDC?
- A. Standard sterilization and disinfection are adequate for bloodborne and emerging pathogens, except prions
- B. Prions need no special consideration because they are viruses
- C. Standard processes are inadequate for HIV and hepatitis B
- D. Prions are destroyed by any low-level disinfectant
Show answer and explanation
Correct answer: A. CDC states that standard sterilization and disinfection procedures are adequate for instruments contaminated with bloodborne or emerging pathogens, with the exception of prions. Follow your facility's CJD policy when prion disease is suspected.
Why the other choices don't fit
- B: Prions are not viruses, and they do need special handling.
- C: CDC says standard procedures are adequate for HIV and HBV.
- D: Prions are the notable exception to standard processes.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 9.a
Question 40 of 50
Sterilization Process · CRCST-040
For each steam sterilization cycle, which set of information does CDC recommend recording?
- A. Only the date
- B. The names of patients who will receive the items
- C. Only the biological indicator result for that week
- D. Sterilizer and cycle type, load identification number, load contents, exposure parameters, operator's name or initials, and the results of mechanical, chemical and biological monitoring
Show answer and explanation
Correct answer: D. CDC lists the type of sterilizer and cycle, load identification number, load contents, exposure parameters such as time and temperature, the operator's name or initials, and monitoring results for each cycle.
Why the other choices don't fit
- A: A date alone can't support recalls or troubleshooting.
- B: Load records track items by load. Patient tracking is a separate step, such as for implants.
- C: Monitoring results are only one part of the record.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 19.e
Question 41 of 50
Departmental Considerations · CRCST-041
CDC gives an example of how long sterilization records (mechanical, chemical and biological) should be kept. What is it?
- A. 30 days
- B. 6 months
- C. 3 years, or longer if statutes or regulations require
- D. Until the sterilizer is replaced
Show answer and explanation
Correct answer: C. CDC recommends retaining sterilization records for a period that complies with standards (giving 3 years as an example), statutes of limitations, and state and federal regulations.
Why the other choices don't fit
- A: Far shorter than CDC's example.
- B: Shorter than CDC's example.
- D: Retention depends on standards and law, not equipment life.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 19.f; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Monitoring"
Question 42 of 50
Preparation and Packaging · CRCST-042
When wrapping a tray in flat sterilization wrap, which problem should you avoid?
- A. Tenting and gapping of the wrap
- B. Attaching an external indicator
- C. Folding the corners
- D. Using two sheets
Show answer and explanation
Correct answer: A. CDC says wrapping should be done in a way that avoids tenting and gapping.
Why the other choices don't fit
- B: External indicators belong on each package.
- C: Folding is part of standard wrapping.
- D: Double wrapping with two sheets is common practice.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Packaging"
Question 43 of 50
Sterile Storage, Transport and Inventory Management · CRCST-043
Under FDA's recall classifications, which class covers a product with a reasonable probability of causing serious adverse health consequences or death?
- A. Class I
- B. Class II
- C. Class III
- D. Market withdrawal
Show answer and explanation
Correct answer: A. FDA defines a Class I recall as a situation where there is a reasonable probability that using the product will cause serious adverse health consequences or death.
Why the other choices don't fit
- B: Class II covers temporary or medically reversible effects, or a remote probability of serious harm.
- C: Class III covers products not likely to cause adverse health consequences.
- D: A market withdrawal involves a minor violation not subject to FDA legal action.
Source: FDA, Recalls Background and Definitions, "Recall classifications"
Question 44 of 50
Patient Care Equipment and Distribution · CRCST-044
A wheelchair was used by a patient on contact precautions, and no dedicated disposable equipment was available. What does CDC recommend before another patient uses it?
- A. Sterilize it
- B. Store it for 72 hours instead of disinfecting
- C. Disinfect it before it is used for another patient
- D. Nothing, because it touched only intact skin
Show answer and explanation
Correct answer: C. CDC recommends that, if dedicated disposable devices aren't available, noncritical patient-care equipment used on a patient on contact precautions be disinfected before use on another patient.
Why the other choices don't fit
- A: Sterilization is for critical items; this is noncritical equipment.
- B: Holding time isn't a disinfection method.
- D: Contact precautions make disinfection between patients necessary.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 4.d
Question 45 of 50
Preparation and Packaging · CRCST-045
Which requirement does CDC set for packaging materials used for sterilization?
- A. Packaging only has to protect the item during transport
- B. Kitchen plastic wrap works for any process
- C. They must be compatible with the sterilization process and have FDA 510(k) clearance
- D. Any material is fine if it is clean
Show answer and explanation
Correct answer: C. CDC recommends packaging that is compatible with the sterilization process, has FDA 510(k) clearance and is strong enough to resist punctures and tears so it keeps out microorganisms and moisture.
Why the other choices don't fit
- A: Packaging must also let sterilant in and keep sterility to the point of use.
- B: Kitchen-type wrap is specifically unacceptable for EO.
- D: Cleanliness alone doesn't make a material suitable.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Recs. 15.a–15.b; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Packaging"
Question 46 of 50
Professional Development and Human Relation Skills · CRCST-046
A coworker from another department asks you to look up a patient's full record so they can see why the patient's surgery was canceled. They are not involved in the patient's care. Which HIPAA concept applies?
- A. Records may be shared if the request is verbal
- B. Minimum necessary: use or disclose only what is needed for a permitted purpose
- C. Anyone in the hospital can see any record
- D. HIPAA applies only to paper records
Show answer and explanation
Correct answer: B. HIPAA's minimum necessary standard requires covered entities to limit uses and disclosures of protected health information to the minimum needed for the intended purpose, with access tied to job duties. Curiosity is not a job duty.
Why the other choices don't fit
- A: How the request is made doesn't change what can be disclosed.
- C: Access is limited by role and need.
- D: The minimum necessary standard is about protected health information generally, not just paper.
Source: HHS Office for Civil Rights, Minimum Necessary Requirement, "How the Rule Works" and "Uses and Disclosures"
Question 47 of 50
Departmental Considerations · CRCST-047
CDC says a central processing area should ideally be divided into at least three areas. Which three?
- A. Decontamination; packaging; and sterilization and storage
- B. Decontamination, sterile storage and the operating room
- C. Receiving, offices and break room
- D. Packaging, case-cart staging and loading dock
Show answer and explanation
Correct answer: A. CDC says central processing ideally has at least three areas (decontamination, packaging, and sterilization and storage), with physical barriers separating decontamination from the other sections to contain contamination on used items.
Why the other choices don't fit
- B: The operating room is a user department, not a processing area.
- C: These are support spaces, not processing areas.
- D: These are workflow spaces, not the three processing areas CDC names.
Source: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), "Sterilizing Practices", "Physical Facilities"
Question 48 of 50
Cleaning, Decontamination and Disinfection · CRCST-048
When should a flexible endoscope be leak tested, and what happens if it fails?
- A. Once a month; if it fails, high-level disinfect it twice
- B. Only when it looks damaged; if it fails, tape the area
- C. Only after sterilization; if it fails, document and continue
- D. As part of each reprocessing cycle; if it fails, remove it from clinical use and have it repaired
Show answer and explanation
Correct answer: D. CDC recommends leak testing each flexible endoscope as part of each reprocessing cycle to detect damage, and removing any scope that fails from clinical use for repair.
Why the other choices don't fit
- A: Monthly testing is too infrequent, and repeating disinfection doesn't fix a leak.
- B: Leaks are often not visible, and tape isn't a repair.
- C: The leak test comes early in reprocessing, and a failed scope can't stay in use.
Source: CDC, Recommendations for Disinfection and Sterilization in Healthcare Facilities, Rec. 7.a
Question 49 of 50
Professional Development and Human Relation Skills · CRCST-049
How many continuing education (CE) credits does HSPA require each year to renew the CRCST?
- A. 6
- B. 12
- C. 24
- D. No CE is required
Show answer and explanation
Correct answer: B. HSPA requires 12 CE credits per year to renew the CRCST. Credits must be technical and relevant to sterile processing. The renewal fee is $50 (or $60 with membership).
Why the other choices don't fit
- A: Lower than HSPA's requirement.
- C: Higher than HSPA's requirement.
- D: HSPA requires annual CE to renew.
Source: HSPA, CRCST Renewal page, "Renewal Requirements"; HSPA, Certification Handbook (revised January 2026), "Maintaining Certification"
Question 50 of 50
Professional Development and Human Relation Skills · CRCST-050
A case cart sheet lists a 'colonoscopy.' What does the word ending -scopy mean?
- A. Creating an opening in
- B. Surgical repair
- C. Removal of
- D. Use of an instrument for viewing
Show answer and explanation
Correct answer: D. MedlinePlus defines -scopy as the use of an instrument for viewing. A colonoscopy is a view of the colon.
Why the other choices don't fit
- A: Creating an opening is -ostomy, as in colostomy.
- B: Surgical repair is not what -scopy means.
- C: Removal of is -ectomy.
Source: MedlinePlus (National Library of Medicine), Understanding Medical Words tutorial (PDF), word-endings section
Score yourself by domain
Count your correct answers in each row. The weights come from HSPA's CRCST Exam Content Outline (revised November 2023). We matched this set to those weights as closely as 50 questions allow.
| CRCST domain | HSPA weight | Questions in this set | Question numbers |
|---|---|---|---|
| 1. Departmental Considerations | 15% | 8 | 3, 9, 12, 18, 24, 30, 41, 47 |
| 2. Cleaning, Decontamination and Disinfection | 21% | 10 | 1, 6, 11, 15, 20, 25, 28, 33, 39, 48 |
| 3. Preparation and Packaging | 21% | 10 | 4, 10, 14, 19, 23, 27, 31, 34, 42, 45 |
| 4. Sterilization Process | 21% | 11 | 2, 5, 8, 13, 16, 21, 26, 29, 32, 36, 40 |
| 5. Sterile Storage, Transport and Inventory Management | 9% | 5 | 7, 17, 22, 35, 43 |
| 6. Patient Care Equipment and Distribution | 5% | 2 | 37, 44 |
| 7. Professional Development and Human Relation Skills | 8% | 4 | 38, 46, 49, 50 |
| Total | 100% | 50 |
Your score here reflects these 50 items only. It isn't a CRCST score and doesn't predict a pass. HSPA doesn't publish a passing percentage; it sets the passing point with a criterion-referenced (Angoff) method and reports pass or fail (HSPA Certification Handbook, "Determination of a Passing Score"). With only 2 to 11 items per domain, treat a weak row as a topic to review, not a measurement.
Domains 2, 3 and 4 carry 21% each, so together they make up 63% of the exam. If you missed several questions in those rows, start there.
What the real CRCST exam looks like
| Detail | Current rule | Source |
|---|---|---|
| Who runs it | Healthcare Sterile Processing Association (HSPA), formerly IAHCSMM | HSPA Certification Handbook, Certification Overview |
| Questions | 150 multiple-choice: 125 scored and 25 unscored pretest questions mixed in at random | Handbook, CRCST "Content & Composition of the Examination" |
| Time | 3 hours, plus an optional 15-minute tutorial that doesn't count against the clock | Handbook, same section |
| Format | Computer-based at a Prometric testing center; you can mark questions and return to them | Handbook, same section and Step 5 |
| Exam fee | $140 | Handbook, CRCST FAQ |
| Scheduling window | 120 days to schedule and sit for the exam after your application is approved | Handbook, CRCST FAQ |
| Results | On-screen pass/fail at the test center. Candidates who don't pass get section results labeled "At or Above," "Below" or "Well Below"; HSPA doesn't give percentage scores | Handbook, "CRCST, CIS, CER & CHL Examination Results" |
| Retake | Wait at least 6 weeks after your last attempt; each retake costs $140 | Handbook, "Re-Testing" |
| Hands-on experience | 400 hours in a sterile processing department, completed before testing (full certification) or within 6 months after passing (provisional) | Handbook, "Requirements for the Examination" |
| Renewal | Every year: 12 technical CE credits and a $50 fee ($60 with membership) | HSPA CRCST Renewal page |
| References HSPA used to write the exam | HSPA's SP Technical Manual, 9th Ed. (2023); ANSI/AAMI ST79 (2017); AORN Guidelines for Perioperative Practice (2025) | Handbook, CRCST "How to Prepare for the Examination" |
The pace on test day works out to 1.2 minutes per question (180 minutes ÷ 150 questions). To practice at that pace, give yourself 60 minutes for these 50.
The 400 hours: full vs. provisional
| Row label | Full certification (HSPA recommends this route) | Provisional certification |
|---|---|---|
| When the hours happen | Before you test | Within 6 months after you pass |
| Who signs | Supervisor documents the hours on the exam application | A manager in the SP department, above you, who witnessed the hours |
| Required breakdown | Decontamination 120 hours; Preparing & Packaging Instruments 120; Sterilization & Disinfection 120; Storage & Distribution 24; Quality Assurance Processes 16 | Same breakdown |
| If time runs short | Not applicable | One 2-month extension, requested by your SP manager before provisional status expires, and only if you are still working or volunteering in an SP department |
| If hours aren't submitted in time | Not applicable | Provisional certification is forfeited and you must pass a retake exam (fees apply) |
Hours can be paid or volunteer and can be split across facilities. HSPA doesn't place candidates, and sites other than hospitals or surgery centers (dental, optical or veterinary clinics, for example) need HSPA approval first. Source: HSPA Certification Handbook, "Requirements for the Examination: Full & Provisional Certification".
Numbers that show up in these questions
These are the specific values the answer explanations rely on, collected in one place. They come from the CDC and OSHA sources cited above. On the job, your department's policies, the device and sterilizer manufacturers' instructions for use (IFU), and ANSI/AAMI ST79 govern. When a value differs, follow the IFU.
| Topic | Value | Source |
|---|---|---|
| Gravity steam, wrapped instruments | 30 min at 250°F (121°C); 15 min at 270°F (132°C) | CDC Table 7 |
| Gravity steam, textile packs | 30 min at 121°C; 25 min at 132°C | CDC Table 7 |
| Prevacuum steam, wrapped instruments | 4 min at 270°F (132°C) | CDC Table 7 |
| Bowie-Dick test | Each day a prevacuum sterilizer is used, before the first load, in an empty chamber | CDC, Steam Sterilization |
| Steam biological indicator | G. stearothermophilus, incubated at 55–60°C; at least weekly, and every load with implants | CDC, Sterilizing Practices |
| After major repair | 3 consecutive empty cycles with BI and CI (plus 3 Bowie-Dick tests in a prevacuum unit) | CDC, Sterilizing Practices |
| EO aeration (PVC tubing example) | 12 hours at 50°C or 8 hours at 60°C | CDC Recommendations, Rec. 14.e |
| EO exposure limits | 1 ppm 8-hour TWA; 0.5 ppm action level; 5 ppm 15-minute excursion limit | CDC, Ethylene Oxide Sterilization |
| Sterile storage clearances | 8–10 in. from the floor; 5 in. from the ceiling (18 in. from sprinkler heads); 2 in. from outside walls | CDC, Sterilizing Practices |
| Sterilization record retention | Per standards and law; CDC's example is 3 years | CDC Recommendations, Rec. 19.f |
| Hepatitis B vaccine offer | Within 10 working days of initial assignment | OSHA 29 CFR 1910.1030(f)(2) |
How to get the most out of this set
- Answer all 50 before opening any explanation. Note how sure you were on each one.
- Score by domain with the table above.
- Read every explanation, including the ones you got right. Right answers you guessed are misses waiting to happen.
- For each miss, open the cited source and read the section around it. Most questions here come from the CDC sterilization guideline, which is free to read.
- Rework the set a week later. Option order on the real exam won't match this page, so make sure you know why the answer is right instead of remembering where it sat.
CRCST practice test FAQ
How many questions are on the CRCST exam?
150 multiple-choice questions in 3 hours. Only 125 count toward your result. The other 25 are unscored pretest questions that look identical, so answer every question carefully.
What score do you need to pass the CRCST?
HSPA doesn't publish a percentage. The passing point is set with the Angoff procedure supplemented by the Beuk compromise method, and your pass or fail is based on the number of scored questions you answer correctly. Be wary of any site that quotes an exact passing percentage.
Are these real CRCST exam questions?
No. Castleport Test Prep wrote every question here from public sources. None are real, recalled or leaked exam items, and the set isn't a simulation of any specific exam form.
Can I take the CRCST before finishing my 400 hours?
Yes. That's the provisional route: pass the exam, then complete and document the 400 hours within 6 months. HSPA recommends finishing, or at least starting, your hours before you test.
What should I study after this practice test?
HSPA names the references used to build the exam: its SP Technical Manual, 9th edition (2023), ANSI/AAMI ST79 (2017) and AORN's Guidelines for Perioperative Practice (2025). HSPA also notes that its Certification Council doesn't require, approve or endorse any specific training materials. For free reading, the CDC sterilization guideline sections cited in the explanations cover much of domains 2, 3 and 4.
How long do I wait to retake the CRCST?
At least 6 weeks from your last exam date. You can submit the retake application sooner, but eligibility starts 6 weeks after the attempt. Each retake costs $140.
Sources
- HSPA, Certification Handbook (revised January 2026): exam composition, fees, eligibility, provisional certification, results, retesting.
- HSPA, CRCST Exam Content Outline (revised November 2023): domains and weights.
- HSPA, CRCST Renewal: CE and renewal fees.
- CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices, Steam Sterilization, Table 7, A Rational Approach, Cleaning, Flash Sterilization, Ethylene Oxide, Summary Recommendations.
- OSHA, 29 CFR 1910.1030 Bloodborne Pathogens; 29 CFR 1910.334; Hazard Communication Standard: Safety Data Sheets (OSHA 3514).
- FDA, Recalls Background and Definitions.
- HHS, HIPAA Minimum Necessary Requirement.
- National Library of Medicine, MedlinePlus Understanding Medical Words tutorial.
- APSIC, CSSD Center of Excellence Audit Checklist; Namibia Ministry of Health and Social Services, CSSD Guideline (October 2023).
Last verified: November 10, 2026. We checked exam format, fees, eligibility, results and retake rules against the HSPA handbook and renewal page, domain weights against the November 2023 content outline, and every answer against the source cited with it.
By the Castleport Test Prep Editorial Team.
Castleport Test Prep is an independent exam prep publisher. We are not affiliated with, endorsed by, or approved by the Healthcare Sterile Processing Association (HSPA). CRCST and other exam and credential names are used only to identify their subjects; trademarks belong to their respective owners. These practice questions are original and unofficial.
No qualified sterile processing specialist review of this version is documented.