Free ARRT CT Practice Test
60 original questions mapped to the ARRT Computed Tomography content specifications effective September 1, 2026, with the answer and a full explanation under each one. These are unofficial practice questions written by Castleport Test Prep, not ARRT exam items.
Show answer and explanation
Answer: C. For a PE study you want peak opacification of the pulmonary arteries, so you trigger off the vessel you're trying to fill — the pulmonary trunk. The SVC (D) enhances first and would trigger the scan too early, before the pulmonary arteries peak, and dense undiluted contrast there can cause streak artifact. The left ventricle (A) and ascending aorta (B) peak after the pulmonary circulation; triggering there times the scan for systemic arteries (a coronary CTA or aortic study), so the pulmonary arteries may already be washing out.
Source: Fleischner Society position paper on CT pulmonary angiography
Source context: The Fleischner position paper linked above addresses pulmonary hypertension. For the pulmonary-trunk bolus-tracking technique in this question, see the published CTPA study's key points, which explicitly describe pulmonary-trunk monitoring as a widely used bolus-tracking approach.
Show answer and explanation
Answer: A. For helical scanning, CTDIvol = CTDIw ÷ pitch, so 12 ÷ 1.5 = 8 mGy. A pitch above 1 spreads the same exposure over more table travel, lowering the average dose in the scanned volume. Multiplying instead of dividing gives 18 mGy (D). C ignores pitch entirely, and B is a distractor with no basis in the formula.
Source: AAPM CT Dose Summit (2011), Assessing Radiation Dose: CTDIvol, DLP and AAPM Report 96 k-factors
Show answer and explanation
Answer: C. The ACR Manual on Contrast Media says patients with eGFR ≥30 and no AKI don't need to discontinue metformin before or after IV iodinated contrast, and don't need routine follow-up renal testing. Holding metformin is reserved for patients with AKI or eGFR below 30. Options A and B describe older, more conservative practice. D withholds a diagnostic test with no indication to do so. Always follow your facility's written policy, which should be built on this guidance.
Source: ACR Manual on Contrast Media, Metformin chapter, p. 48
Show answer and explanation
Answer: C. Pitch = table travel per rotation ÷ total beam collimation. Total collimation is 64 × 0.625 mm = 40 mm, so 52 ÷ 40 = 1.3. Choice A comes from dividing the wrong way (40 ÷ 52). A pitch of 1.0 (B) would mean the table moved exactly 40 mm per rotation. Always calculate total collimation first — not the single-row width.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: B. Adenocarcinoma usually enhances less than normal pancreas. Imaging during the pancreatic parenchymal phase maximizes the attenuation difference between enhancing normal pancreas and the typically hypoenhancing tumor. The portal venous phase that follows is useful for liver metastases, nodes and venous anatomy. A noncontrast phase (A) cannot show differential enhancement; excretory imaging (C) is a urinary-tract phase; and delayed equilibrium imaging (D) is not the phase optimized for primary tumor conspicuity.
Source: AJR: pancreatic-phase versus portal-vein-phase CT of the pancreas
Show answer and explanation
Answer: D. The window is centered on the level and extends half the width in each direction: 40 − 200 = −160 and 40 + 200 = +240. Anything above +240 HU displays white and anything below −160 HU displays black. C forgets to center on the level of 40. A starts at zero, and B adds the full width to each side instead of half.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: A. Fresh clotted blood is denser than brain parenchyma and appears bright on noncontrast CT, which is why noncontrast head CT is the first study when hemorrhage is suspected. Blood becomes isoattenuating (C) and then hypoattenuating (B) as it ages over days to weeks, so timing matters. Fat density (D) doesn't describe hemorrhage. Giving IV contrast first can confuse the picture — enhancing structures can mimic or hide blood.
Show answer and explanation
Answer: B. Hypotension with bradycardia points to a vasovagal reaction. The ACR adult reaction card manages it with oxygen, leg elevation, and a wide-open normal saline bolus, adding atropine if hypotension is refractory. Anaphylaxis typically produces hypotension with tachycardia — that is when epinephrine (A) is indicated. Diphenhydramine (C) can worsen hypotension, and an inhaler (D) treats bronchospasm, which this patient doesn't have.
Source: ACR adult contrast reaction card (updated Sept. 2026)
Show answer and explanation
Answer: D. In multidetector CT, the acquired detector width sets the minimum reconstructed slice thickness. Data collected in 1.25 mm channels can be reconstructed at 1.25 mm or thicker, never thinner. Raw data (A) lets you change the kernel, field of view and thickness only within that limit. A sharper kernel (C) changes edge sharpening, not slice thickness, and reformatting (B) can't create z-axis detail that wasn't acquired. Choose thin collimation up front when thin images may be needed.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: A. Sublingual nitroglycerin is used to dilate the coronary arteries for coronary CTA, but it is contraindicated after a recent phosphodiesterase-5 inhibitor such as sildenafil because the combination can cause severe hypotension. The contrast (B) is still needed. The heart rate is already at the usual goal of 60 bpm or less, so no more beta-blocker is needed, but metoprolol itself isn't the interacting drug (C). Proceeding unchanged (D) ignores a dangerous interaction. Notify the supervising physician and document.
Show answer and explanation
Answer: A. Rings are an equipment artifact of rotate-rotate (third-generation) scanners: one detector element reading incorrectly traces a circle as the gantry rotates. Recalibration (an air calibration) or service is the fix. Motion (B) makes streaks, blurring or doubled edges. Beam hardening (C) causes dark streaks between dense structures and cupping. Partial volume averaging (D) blurs small structures and alters their CT numbers. None of those produce perfect circles.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: A. Stones are easiest to see without contrast, and a low-dose technique is usually adequate because stones are very high-contrast against surrounding tissue. Excreted contrast in the collecting system (C) can hide a stone that's just as dense. Portal venous imaging (B) and renal CTA (D) answer other questions. A urogram or added phases may follow if the radiologist needs them.
Source: ACR Appropriateness Criteria — Acute Onset Flank Pain, Suspicion of Stone Disease
Show answer and explanation
Answer: C. In soft tissue at the effective energies of a CT beam, Compton scattering dominates. The photoelectric effect (A) is more important at lower energies and in high-atomic-number materials like iodine, calcium and bone, which is why they look so bright. Coherent scattering (B) is a small contributor. Pair production (D) requires photon energies above 1.02 MeV, far beyond diagnostic CT.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: D. Until the spine is cleared, the neck must not be moved. A thin-section helical acquisition with immobilization in place produces high-quality sagittal and coronal reformats, without repositioning the patient. Removing the collar or flexing the neck (A) risks spinal cord injury. Thick 5 mm data (C) produces poor reformats that can miss nondisplaced fractures. Gantry tilt (B) can't produce a true coronal plane, and it doesn't replace reformats.
Source: ACR CT Accreditation — Adult Spine technique parameters
Show answer and explanation
Answer: D. Changing the reconstruction kernel, the display field of view, or the reconstructed slice thickness requires the raw projection data, a process called retrospective reconstruction. Images already on PACS (A) are finished image data; you can window or reformat them, but you can't change their kernel. A new scout (C) does nothing here, and as long as raw data is still stored on the scanner, rescanning the patient (B) adds unnecessary dose.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: C. The ACR recommends elevating the affected extremity and permits either warm or cold compresses; a health care provider should examine the limb, the event should be documented, and outpatients need clear instructions about when to seek care because severe injury can evolve after discharge. Surgical consultation is driven by concerning signs such as increasing pain, altered perfusion or sensation, reduced range of motion, or blistering—not by an extravasated volume alone (A). Routine aspiration through the IV catheter (B) and hyaluronidase injection (D) are not recommended.
Source: ACR Manual on Contrast Media — Extravasation of Contrast Media
Show answer and explanation
Answer: D. Retrospective gating acquires data continuously while the ECG is recorded, so images can be reconstructed at any point in the R-R interval — necessary for functional analysis across the cycle. The tradeoff is higher dose (which ECG-based tube current modulation can reduce). Prospective triggering (A) exposes only a chosen phase window, lowering dose but limiting functional data. Non-gated scans (C) suffer cardiac motion, and calcium scoring (B) is a single noncontrast phase.
Show answer and explanation
Answer: A. Pixel size = DFOV ÷ matrix size = 250 mm ÷ 512 ≈ 0.49 mm. Shrinking the DFOV or enlarging the matrix makes pixels smaller and can improve displayed spatial resolution, up to the limits set by the acquisition. D is what you'd get with a 500 mm DFOV; B and C come from misplaced decimals or dividing the wrong way.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: C. Absolute washout = (enhanced − delayed) ÷ (enhanced − unenhanced) × 100 = (100 − 40) ÷ (100 − 20) = 60 ÷ 80 = 75%. An absolute washout of 60% or more is the commonly used threshold supporting an adenoma. (Relative washout here — (100 − 40) ÷ 100 = 60% — also exceeds its usual 40% threshold.) B uses the relative-washout figure and mislabels it. A and D come from arithmetic errors, and washout alone doesn't diagnose pheochromocytoma.
Show answer and explanation
Answer: C. First DLP = CTDIvol × scan length = 10 mGy × 35 cm = 350 mGy·cm. Then effective dose ≈ DLP × k = 350 × 0.014 = 4.9 mSv. D confuses DLP with effective dose (and uses the wrong unit); A multiplies only CTDIvol by k; B skips a step. Remember effective dose is a population-level estimate for comparing exams, not a precise dose to the individual patient.
Source: AAPM CT Dose Summit (2011), Assessing Radiation Dose: CTDIvol, DLP and AAPM Report 96 k-factors
Show answer and explanation
Answer: C. Current ACR adult temporal-bone accreditation technique calls for axial and coronal images no thicker than 2 mm, a sharp/high-spatial-frequency reconstruction, and a small display field of view. Those choices preserve fine bony detail in structures such as the ossicles and otic capsule. Thick, smooth or soft-tissue reconstructions (A, B) sacrifice spatial resolution. Contrast-enhanced brain-algorithm images (D) are not a substitute for the dedicated high-resolution bony reconstruction.
Source: ACR CT Accreditation — Adult Head and Neck technique parameters
Show answer and explanation
Answer: D. Iterative (and model-based) reconstruction repeatedly compares and corrects estimated images against the measured data and models noise, producing less noise than filtered back projection at the same dose. Protocols can then use lower mAs or kVp. It is a reconstruction method, so it doesn't change rotation time (A) or detector hardware (B), and it still needs raw data (C). Heavy strength settings can change image texture, which is why radiologists set the levels.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: B. The ACR treats eGFR below 30 (in patients not on dialysis) as the group at potential risk for contrast-induced kidney injury from IV iodinated contrast, so a radiologist should weigh the indication and alternatives before contrast is given. For these patients, ACR–National Kidney Foundation consensus statements recommend IV normal saline prophylaxis unless it's contraindicated, such as in heart failure. A ignores that threshold. Changing dose or rate on your own (C) or switching agents without an order (D) are outside the technologist's role, and they don't resolve the risk decision.
Source: ACR–NKF consensus statements on IV iodinated contrast in kidney disease, Radiology (2020)
Source: ACR Manual on Contrast Media, Post-Contrast AKI chapter, p. 39
Show answer and explanation
Answer: C. The USPSTF recommends annual low-dose CT for adults aged 50 to 80 with at least a 20 pack-year history who currently smoke or quit within the past 15 years. Only C meets all three criteria. A is under 50, B quit more than 15 years ago, and D is over 80. Insurer and program criteria can differ, so screening programs verify eligibility for each patient.
Show answer and explanation
Answer: A. As the polychromatic beam passes through dense bone, low-energy photons are preferentially absorbed. The beam becomes harder, the scanner underestimates attenuation, and dark streaks appear between dense structures — classically between the petrous ridges. Thin slices, bone-correction software and good positioning reduce it. Rings (B) are circular, out-of-field artifacts (C) occur at the edge when anatomy extends beyond the scan field, and tube arcing (D) is an equipment fault producing streaks across the image.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: A. HCC gets its blood supply mainly from the hepatic artery, so it enhances avidly when hepatic arterial enhancement peaks, before the background liver (supplied mostly by the portal vein) has fully enhanced. Portal venous (B) and delayed (D) phases show the "washout" and capsule features. The noncontrast phase (C) is a baseline for judging enhancement. Imaging too early (early arterial) can miss hyperenhancement, which is why bolus tracking is used for timing.
Source: ACR LI-RADS support: technique (late arterial phase)
Show answer and explanation
Answer: B. A MIP projects the highest-attenuation voxel along each ray, so dense contrast-filled vessels stand out from surrounding tissue in an angiogram-like image. Dense calcification and bone are also bright on MIP, which can hide the lumen, so they're often reviewed along with curved reformats or volume rendering. MinIP (A) highlights low-attenuation structures like airways. A single thin MPR (C) can't show a tortuous vessel's full course, and skin-surface display (D) doesn't show vessels.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: A. Iodine's attenuation rises sharply just above its K-edge at 33.2 keV. A lower-kVp beam has more photons near that energy, so iodine enhancement increases. That lets the protocol keep vascular contrast-to-noise at a lower dose, or reduce contrast volume. The tradeoff is more noise, because fewer photons penetrate (B is backward), so it fits smaller patients best. Lower kVp doesn't eliminate beam hardening (C) — it can worsen it — and it has nothing to do with temporal resolution (D).
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: C. The orbital floor runs roughly horizontally, so axial slices (A) cut parallel to it and can miss a depressed fragment or herniated orbital fat or muscle. Coronal images — usually reformatted from thin axial data — cut across the floor and show depression directly. A midline sagittal plane (B) misses both orbits; a parasagittal reformat through each orbit can help. An oblique optic-nerve plane (D) is for the nerve, not the floor.
Source: ACR Appropriateness Criteria — Orbital Imaging and Vision Loss
Show answer and explanation
Answer: A. The ACR Manual lists a 13-hour oral regimen: prednisone 50 mg at 13, 7 and 1 hour before contrast, with diphenhydramine 50 mg 1 hour before (the antihistamine is considered optional). Another listed option is methylprednisolone 32 mg at 12 and 2 hours before. The ACR notes that regimens under 4–5 hours haven't been shown to work, which rules out B and D, and an antihistamine alone (C) isn't a corticosteroid premedication regimen. Breakthrough reactions can still occur, so reaction readiness stays in place.
Source: ACR Manual on Contrast Media, Patient Selection chapter, premedication regimens, pp. 9–10
Show answer and explanation
Answer: A. Each voxel shows the average attenuation of everything inside it. A 4 mm nodule in a 5 mm thick voxel is averaged with surrounding air-filled lung, lowering its CT number and making it faint. Thinner slices reduce partial volume averaging. Beam hardening (B) creates dark streaks near dense objects; noise (C) makes graininess rather than a consistent lower value; out-of-field (D) affects the image edges.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: A. TAVR planning needs precise annulus and aortic-root measurements (ECG-gated, to freeze motion) and an assessment of the access route — the aorta and iliofemoral arteries — for size, tortuosity and calcification. Noncontrast chest (B) can't measure lumens. Intracranial CTA (C) and a venous-phase abdomen (D) don't answer the valve or access questions.
Source: SCCT expert consensus on CT imaging before TAVI/TAVR, JCCT (2019)
Show answer and explanation
Answer: D. Slip rings transfer power and data across the rotating gantry through sliding contacts, so the gantry can spin continuously in one direction. That made helical (spiral) scanning possible. Analog-to-digital conversion (A) happens in the data acquisition system. Filtration (C) and beam shaping (B) are done by filters, including the bowtie filter, and collimators.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: B. A bladder tear can stay hidden if the bladder isn't under distending pressure. Retrograde filling with dilute iodinated contrast through a catheter fills the bladder adequately, so leakage into the perivesical space (extraperitoneal) or around bowel loops (intraperitoneal) becomes visible. A is the reverse of the truth. C is wrong because contrast is diluted for CT to avoid streak artifact. D is wrong because retrograde filling requires a catheter.
Source: ACR Practice Parameter for Imaging of the Urinary Tract in Adults
Show answer and explanation
Answer: C. CTDIvol is inversely proportional to pitch: 1 ÷ 1.375 ≈ 0.73, a drop of about 27%. A reverses the relationship. Note that some scanners use an "effective mAs" setting that automatically raises tube current as pitch increases to keep CTDIvol constant, which is why the stem specifies constant tube current. Higher pitch also slightly reduces z-axis resolution.
Source: AAPM CT Dose Summit (2011), Assessing Radiation Dose: CTDIvol, DLP and AAPM Report 96 k-factors
Show answer and explanation
Answer: B. Only nonionic iodinated agents approved and labeled for intrathecal use, at labeled concentrations and doses, are used for myelography. Inadvertent intrathecal injection of an ionic agent (A) can cause severe, potentially fatal neurotoxic reactions — which is why intrathecal contrast is verified before injection. Barium (C) is never injected into the thecal sac. Gadolinium (D) isn't a routine intrathecal agent.
Source: DailyMed — iohexol injection prescribing information
Show answer and explanation
Answer: D. For clinically significant venous air embolism, the ACR describes giving 100% oxygen and placing the patient left side down, which helps keep air in the right side of the heart away from the pulmonary outflow tract. Hyperbaric oxygen may be considered, and CPR started if arrest occurs. A puts the wrong side down without oxygen. C treats an allergic reaction, not air. B continues the source of the problem. Purging air from the syringe and tubing before every injection is the prevention.
Source: ACR Manual on Contrast Media, Safe Injection chapter, pp. 14–15
Show answer and explanation
Answer: B. Agatston scoring uses a noncontrast, ECG-synchronized scan. Calcified lesions at or above the 130 HU threshold are measured, and each lesion's area is weighted by a density factor based on its peak attenuation. Iodinated contrast (A) would be indistinguishable from calcium. A 0 HU threshold (C) would count water-density tissue. DLP (D) is a dose metric, not a calcium measurement.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: C. By definition water is 0 HU (air is −1000 HU), so a water reading this far from 0 is a calibration problem that will shift every CT number, affecting things like lesion characterization. Follow the facility's QC procedure, which typically calls for recalibration and retesting, then service if it fails, before scanning patients. A and B accept an out-of-tolerance result. Changing the display window (D) doesn't change measured CT numbers.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: D. In the supine position, gravity-dependent density in the posterior lungs can mimic early fibrosis. If it disappears when the patient is prone, it was dependent atelectasis; if it persists, it's real disease. Prone images add dose rather than reduce it (A), HRCT is typically noncontrast (C), and thin, sharp-kernel sections remain essential (B). Expiratory images are another common add-on, used to assess air trapping.
Show answer and explanation
Answer: A. An unenhanced attenuation of 10 HU or less is the widely used threshold for a lipid-rich adenoma: intracellular fat lowers the CT number. Acute hemorrhage (C) is typically higher density. Pheochromocytoma (B) and metastases (D) usually measure above 10 HU unenhanced, so a nodule above that threshold may need washout imaging or other evaluation.
Show answer and explanation
Answer: B. Helical reconstruction interpolates data from either side of each image location, so the scanner must expose a bit beyond the planned range at both ends. Over-ranging matters most for short scan lengths and wide beams. Adaptive (dynamic) collimation that blocks the unused part of the beam at the start and end reduces it. Patient motion (A), dose alerts (C) and field-of-view choices (D) are different issues.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: D. Perfusion maps (cerebral blood flow, cerebral blood volume, and transit-time measures such as MTT or Tmax) estimate how much tissue is already infarcted (core) versus at risk but salvageable (penumbra), which informs treatment decisions. Hemorrhage (A) is excluded on noncontrast CT, which still comes first (B). Perfusion CT doesn't measure CSF flow (C).
Source: ACR Appropriateness Criteria — Cerebrovascular Diseases: Stroke and Stroke-Related Conditions
Show answer and explanation
Answer: A. A reconstruction interval smaller than the slice thickness produces overlapping images. That better samples small structures along the z-axis and gives smoother, more accurate MPR and 3D images. Overlap is a reconstruction choice made after the exposure, so it doesn't change dose (B), temporal resolution (C) or scan time (D) — it does create more images to store.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: B. A few scattered hives with normal vital signs and no respiratory symptoms are a mild allergic-like reaction. Current ACR guidance allows observation alone for limited urticaria; an antihistamine may be considered if symptoms are bothersome or progress. Epinephrine (A) is reserved for a more severe reaction such as anaphylaxis with respiratory compromise or hypotension. Immediate discharge (C) without observation is inappropriate, and a wide-open fluid bolus (D) is not indicated when blood pressure is normal.
Source: ACR adult contrast reaction card (updated Sept. 2026)
Show answer and explanation
Answer: C. Temporal resolution is how well a scanner freezes motion, and it depends mainly on how quickly the projection data for one image is acquired. Faster rotation — plus techniques such as dual-source acquisition and partial-scan or multisegment reconstruction — improves it, which matters for cardiac imaging. Matrix size (A) and slice thickness (B) affect spatial resolution; kVp (D) affects penetration, contrast and noise.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: D. When perforation is suspected, water-soluble iodinated contrast is used first because barium (A) that leaks into the mediastinum can cause serious inflammation and persists in the tissues. Gadolinium (C) isn't an oral CT agent. Air alone (B) can't show a leak tract the way positive contrast can. Aspiration risk is also part of contrast selection, so follow the radiologist's protocol.
Show answer and explanation
Answer: B. The excretory phase is timed for contrast to fill the collecting systems, ureters and bladder, so filling defects such as urothelial tumors and other abnormalities stand out against dense urine. Some protocols use a split-bolus technique to combine phases. Stones (C) are assessed on the noncontrast phase. Arterial or nephrographic phases (A) characterize renal masses. An extra phase adds dose rather than reducing it (D).
Show answer and explanation
Answer: D. The DAS sits between the detectors and the computer. It samples and amplifies detector output, digitizes it with analog-to-digital converters, and transmits the raw data for reconstruction. The generator (A) supplies tube voltage, PACS (C) stores finished images, and the table (B) is controlled separately.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: B. Dose-check features compare planned exposure with facility-set values before scanning. A notification is a per-series prompt to double-check the protocol. An alert looks at the accumulated dose at any location across the whole exam and calls for stricter review: per AAPM guidance, the operator enters their name and confirms or changes the settings, and a facility may also require a password. C and D describe things these tools don't do, and A ignores the distinction.
Source: AAPM, Recommendations Regarding Notification and Alert Values for CT Scanners (NEMA XR 25), 2011
Show answer and explanation
Answer: B. 3D volume-rendered images let the surgeon see the fracture pattern and fragment displacement as a whole, and digitally removing the femoral head exposes the acetabular articular surface. MinIP (A) is for airways and low-density structures, curved vascular reformats (C) are for vessels, and perfusion maps (D) are for blood flow.
Show answer and explanation
Answer: C. The ACR advises power injecting into a central venous catheter or PICC only when it is certified for power injection, and then within the manufacturer's flow-rate limit. An unverified line risks catheter rupture or tip migration. A and B assume a rating you haven't confirmed. Forceful hand injection (D) can also generate high pressure in a small-bore catheter.
Source: ACR Manual on Contrast Media, Safe Injection chapter, pp. 14–15
Source: ACR Manual on Contrast Media, Extravasation chapter, pp. 17–24
Show answer and explanation
Answer: D. Acquiring data at two energy spectra lets software separate materials by how their attenuation changes with energy. It can then subtract iodine to create virtual noncontrast images, or create iodine maps and virtual monoenergetic images. MPR (A), retrospective kernel changes (C) and bolus tracking (B) are available on single-energy scanners.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: B. IV contrast helps distinguish enhancing vessels from lymph nodes and improves depiction of inflammatory collections and masses, including peripheral enhancement around an abscess. Current ACR guidance rates contrast-enhanced CT neck as appropriate for common neck-mass/infectious evaluations, and ARRT's current clinical-experience documents include contrast-enhanced soft-tissue neck CT. Contrast does not lower radiation dose (A), it is not given to opacify the esophagus (C), and the airway is visible without IV contrast (D).
Source: ACR Appropriateness Criteria — Neck Mass/Adenopathy
Source: ARRT, Updated CT clinical experience requirements (May 29, 2026)
Show answer and explanation
Answer: A. DICOM (Digital Imaging and Communications in Medicine) defines the medical image format and the network protocols that let scanners, PACS and workstations from different vendors exchange images and their metadata. HL7 (B) carries text-based clinical messages such as orders and reports between information systems. JPEG (C) is a general image compression format that lacks DICOM's medical metadata. HIPAA (D) is a privacy law, not an image standard.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: D. Changing position moves residual fluid and stool and shifts where the colon distends best. A polyp hidden under fluid supine, or a segment collapsed in one position, is often seen in the other. Distension is achieved with insufflated gas, usually CO2. Screening CTC uses low-dose technique (A is backward), routine screening doesn't require IV contrast (C), and it doesn't measure perfusion (B).
Show answer and explanation
Answer: B. Current ACR adult hip CT technique requires axial images with sagittal, coronal, and long-oblique-axis reformations. Axial-only (A) or coronal-only (C) imaging omits required planes. Three-dimensional renderings (D) may be useful for selected problems, but they do not replace the diagnostic cross-sectional and multiplanar image set.
Source: ACR CT Accreditation — Adult MSK technique parameters
Show answer and explanation
Answer: C. Bowtie filters are shaped to deliver more radiation to the center of the field than to the edges, assuming the patient is centered. When the patient is off-center, parts of the body receive the less-filtered, higher-intensity portion of the beam, raising surface dose, while other areas can be noisier. Off-center positioning also changes the patient's apparent size on the localizer, which can make automatic exposure control set inappropriate tube current. Centering at isocenter is a basic dose-optimization step.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: D. Perfusion needs time–attenuation curves, so the same tissue is scanned many times over roughly a minute. Cine mode keeps the table stationary while the gantry repeatedly images one volume; shuttle (toggling table) mode moves back and forth to cover more length. A single helical pass (A) or one pass of contiguous slabs (C) gives one time point. A scout (B) isn't a cross-sectional acquisition.
Source: AAPM Diagnostic Radiology Residents Physics Curriculum, Module 11: Computed Tomography
Show answer and explanation
Answer: B. Current ACR adult knee CT technique requires axial images with sagittal and coronal reformations. Coronal-only (A) or sagittal-only (C) imaging is incomplete. Three-dimensional rendering (D) can supplement selected studies but does not replace the required diagnostic cross-sectional and multiplanar images.
Source: ACR CT Accreditation — Adult MSK technique parameters
Answer key
| # | Answer | # | Answer | # | Answer | # | Answer |
|---|---|---|---|---|---|---|---|
| 1 | C | 16 | C | 31 | A | 46 | C |
| 2 | A | 17 | D | 32 | A | 47 | D |
| 3 | C | 18 | A | 33 | D | 48 | B |
| 4 | C | 19 | C | 34 | B | 49 | D |
| 5 | B | 20 | C | 35 | C | 50 | B |
| 6 | D | 21 | C | 36 | B | 51 | B |
| 7 | A | 22 | D | 37 | D | 52 | C |
| 8 | B | 23 | B | 38 | B | 53 | D |
| 9 | D | 24 | C | 39 | C | 54 | B |
| 10 | A | 25 | A | 40 | D | 55 | A |
| 11 | A | 26 | A | 41 | A | 56 | D |
| 12 | A | 27 | B | 42 | B | 57 | B |
| 13 | C | 28 | A | 43 | D | 58 | C |
| 14 | D | 29 | C | 44 | A | 59 | D |
| 15 | D | 30 | A | 45 | B | 60 | B |
Review your misses by section
Compare your accuracy within each row rather than using raw miss counts alone, because this practice set has different numbers of questions in each section. Start with your lowest section percentage; if two sections are close, use the last column as a secondary priority because it shows how many scored questions ARRT assigns to that section.
| Section | Question numbers | In this set | Scored on the real exam |
|---|---|---|---|
| Patient Care — Patient Interactions and Management | 3, 8, 16, 23, 30, 37, 45, 52 | 8 | 21 |
| Safety — Radiation Safety and Dose | 2, 13, 20, 28, 35, 42, 50, 58 | 8 | 21 |
| Image Production — Image Formation | 4, 9, 15, 22, 27, 33, 38, 44, 49, 53, 59 | 11 | 30 |
| Image Production — Image Evaluation and Archiving | 6, 11, 18, 25, 31, 39, 46, 55 | 8 | 22 |
| Procedures — Head, Spine, and Musculoskeletal | 7, 14, 21, 29, 36, 43, 51, 57, 60 | 9 | 25 |
| Procedures — Neck and Chest | 1, 10, 17, 24, 32, 40, 47, 54 | 8 | 22 |
| Procedures — Abdomen and Pelvis | 5, 12, 19, 26, 34, 41, 48, 56 | 8 | 24 |
| Total | 60 | 165 |
What your score means
Your result here is the share of these 60 questions you got right, nothing more. If you answer 45 of 60 correctly, that's 75% on this set. It is not an ARRT scaled score of 75, and it doesn't predict whether you'll pass.
ARRT reports results on a scaled score from 1 to 99, and you need a 75 to pass (2026 Postprimary Handbook, pp. 56–57). A scaled score adjusts for differences in difficulty between exam versions, so it isn't a percentage. A section with only 8 questions here is a rough signal, not a precise measure — use it to decide what to study, not to grade yourself.
How to use this practice test
- Take it cold. Answer every question before reading the explanation, and don't look anything up.
- Read every explanation, including ones you got right by guessing. The wrong options teach as much as the right one.
- Study your weakest section using the explanations and the ARRT content specifications for that section.
- Retake only your misses after a few days of review, so you're testing what you learned rather than what you remember from an hour ago.
The set covers all seven ARRT subcategories roughly in proportion to the real exam, including topics new in September 2026: power injection and PICC lines (Q52), air embolism (Q37), dose alerts (Q50), cardiac CT and TAVR (Q10, Q17, Q32), esophagram (Q47), urogram (Q48), renal stone (Q12), retrograde cystogram (Q34), CT colonography (Q56) and surgical planning (Q51). All 60 are single-best-answer questions without images. ARRT's exam is mostly in that format, but it can also include multiple-answer, sorting, image-selection and short-video questions (Handbook, p. 39).
What's on the ARRT CT exam now
| Current ARRT rule | |
|---|---|
| Who takes it | Technologists on ARRT's postprimary pathway who hold a supporting credential in Radiography (ARRT), Nuclear Medicine Technology (ARRT or NMTCB) or Radiation Therapy (ARRT), plus CT structured education and clinical experience (ARRT supporting categories) |
| Questions | 195 total: 165 scored + 30 unscored pilot questions you can't identify (Handbook, p. 46) |
| Test time | 195 minutes, within a 215-minute appointment that includes the tutorial, nondisclosure agreement and survey (Handbook, p. 46) |
| Passing | Scaled score of 75 on a 1–99 scale (Handbook, pp. 56–57) |
| Application fee | $225; $450 if NMTCB is your supporting credential (ARRT fees) |
| Exam window | 365 days to schedule and test at a Pearson VUE center (Handbook, p. 32) |
| Attempts | 3 attempts within 3 years of your first exam window opening; each needs a new application and fee (Handbook, p. 18) |
Exam facts last verified October 27, 2026.
What changed on September 1, 2026
ARRT moved two scored questions into Image Formation and shifted one from Abdomen and Pelvis to Neck and Chest. The total stayed at 165 (content specifications, p. 1; prior counts from ARRT Annual Exam Report 2025, p. 23).
| Subcategory | Through Aug. 31, 2026 | From Sept. 1, 2026 |
|---|---|---|
| Patient Interactions and Management | 22 | 21 |
| Radiation Safety and Dose | 22 | 21 |
| Image Formation | 28 | 30 |
| Image Evaluation and Archiving | 22 | 22 |
| Head, Spine, and Musculoskeletal | 25 | 25 |
| Neck and Chest | 21 | 22 |
| Abdomen and Pelvis | 25 | 24 |
| Total scored | 165 | 165 |
ARRT also added time-out, power-injection options and scan-timing techniques to Patient Care, adverse event reporting to Safety, and external equipment to Image Production. In Procedures, it added esophagram, urogram/IVU, renal stone, retrograde cystogram and CT colonography, expanded cardiac CT, changed the pediatric focus to infants and young children, and added surgical planning (ARRT, Oct. 31, 2025). If you studied from an older outline or question bank, check these topics first.
Did the passing standard change too?
Yes. For anyone testing on or after September 1, 2026, ARRT applies an updated CT passing standard (ARRT, Aug. 27, 2026). You still need a scaled 75, but it takes more correct answers to get there. On the exam form ARRT used to set the new standard, 75 now equals 112 of 165 scored questions correct, up from about 110 on that same form. The exact number shifts from form to form, because forms differ in difficulty (Standard Setting Report, p. 6).
ARRT expects the first-time pass rate to settle around 69% under the new standard (Standard Setting Report, p. 6). In 2025, under the earlier standard, 74.7% of 4,966 first-time CT candidates passed (Annual Exam Report 2025, p. 23).
Frequently asked questions
Are these real ARRT CT exam questions?
No. All 60 are original questions we wrote from ARRT's published content outline, with the teaching points checked against the clinical and technical sources linked in the explanations. ARRT's exam questions are confidential, and ARRT tells candidates to be careful with study materials that claim to contain ARRT test questions (Handbook, p. 40).
Does ARRT offer an official CT practice test?
ARRT does not publish or endorse a CT mock practice test in the current handbook. It says it doesn't recommend or endorse review programs, mock registries or study guides, and it points candidates to the content specifications instead (Handbook, p. 40). At the test center, an optional tutorial lets you try the question screens and on-screen calculator before the exam starts.
What happens if I don't pass?
You have three attempts within three years, and each one needs a new application and fee. Before retesting, compare the current content specifications with the version you studied, because ARRT updates them (Handbook, pp. 18, 60). The ARRT exam retake policy page walks through the process.
Next steps
- Applying and scheduling: ARRT exam registration
- After test day: ARRT exam results
- Keeping the credential: ARRT continuing education requirements
- Official requirements and documents: ARRT Computed Tomography credential page
Sources
- ARRT, Computed Tomography Examination Content Specifications, effective Sept. 1, 2026 (Board approved July 2025)
- ARRT, 2026 Postprimary Eligibility Pathway Handbook
- ARRT, Supporting category requirements and Application fees
- ARRT, Updated CT documents (Oct. 31, 2025) and Updated CT clinical experience requirements (May 29, 2026)
- ARRT, Updated CT passing standard (Aug. 27, 2026) and Standard Setting Report: Computed Tomography, effective September 2026
- ARRT, Annual Exam Report 2025
- American College of Radiology, ACR Manual on Contrast Media, adult contrast reaction card, esophagram and upper GI practice parameter (2024), LI-RADS technique guidance and incidental adrenal mass white paper (2017)
- ACR and National Kidney Foundation, consensus statements on IV iodinated contrast in kidney disease (2020)
- AAPM, notification and alert values for CT dose check (2011) and CT dose metrics, 2011 CT Dose Summit
- Society of Cardiovascular Computed Tomography, CCTA acquisition guidelines (2016) and CT before TAVI/TAVR consensus (2019)
- U.S. Preventive Services Task Force, Lung Cancer: Screening (2021)
Castleport Test Prep is an independent exam prep publisher. It is not affiliated with, endorsed by, or approved by The American Registry of Radiologic Technologists (ARRT). ARRT and other exam and credential names are used only to identify the exam this resource covers; trademarks belong to their respective owners. These are original, unofficial practice questions, not ARRT exam items. This page is a study resource and doesn't replace your facility's protocols, your supervising radiologist's direction, or ARRT's official requirements.