Free NBRC PFT Practice Test: 50 Questions With Explanations
Fifty original, unofficial practice questions for the NBRC Pulmonary Function Technology (PFT) exam, the one exam behind both the CPFT and RPFT credentials. Pick an answer, then open the explanation under it; this set samples the October 2022 PFT content outline, with sources beside each answer.
Practice test: 50 questions
Part 1 — Instrumentation and equipment (Q1–16)
Q1. Calibration verification
Before the first patient, a technologist runs the daily calibration verification on a flow-sensor spirometer with a 3.00-L syringe. Strokes at low, medium and high flow read 3.10, 3.11 and 3.12 L. What should happen next?
Show answer
Answer: C. The 2019 ATS/ERS spirometry update allows ±3% (±2.5% for the spirometer plus ±0.5% for the syringe). For a 3.00-L syringe that is 2.91–3.09 L, and all three strokes sit above it. A failed check means you don't test patients until the cause is fixed and the verification passes. Likely causes include a damp or contaminated sensor, a syringe leak, or wrong ambient settings. There is no ±5% allowance (A). Correcting results after the fact (B) doesn't replace a passing check. Overestimated volumes aren't "conservative" (D); they can hide a genuinely low FVC. Source: 2019 ATS/ERS spirometry standard, Calibration verification.
Q2. Plethysmograph calibration
On a constant-volume body plethysmograph, which signals should be physically calibrated daily, or before use?
Show answer
Answer: A. The 2023 lung volume standard calls for daily physical calibration of the mouth-pressure transducer and the box signal. Calibrate more often when room temperature or pressure is changing. The isothermal lung analog check is a separate monthly accuracy validation, not a substitute for daily calibration (B, C). Biological controls supplement calibration; they don't replace it (D). Source: 2023 ERS/ATS lung volumes standard, Quality control.
Q3. DLCO volume check
On a single-breath DLCO system, the daily 3-L syringe volume check reads 3.08 L. Your lab's spirometry tolerance is ±3%. What is the correct call for DLCO testing?
Show answer
Answer: B. The 2017 ERS/ATS DLCO standard sets a daily 3-L syringe check within 2.5%. An error of 0.08 L on 3.00 L is 2.7%, so this reading is within the spirometry tolerance but outside the DLCO tolerance. Volume error isn't harmless (C): DLCO is the product of KCO and VA, so an error in VA carries straight into DLCO. The 0.5% figure (D) is the accuracy of the calibration syringe itself, not the instrument's tolerance. Sources: 2017 ERS/ATS DLCO standard, Equipment calibration and quality control, item 2 (PDF p. 7); 2022 ERS/ATS interpretation standard, Gas transfer impairments (PDF p. 23; DLCO = KCO × VA).
Q4. In-line filters
Your lab starts using disposable in-line bacterial/viral filters for every spirometry patient. How should daily calibration verification be done?
Show answer
Answer: D. The 2019 update says that if a filter is used for testing, it must also be in place for calibration and verification, so the check reflects the system patients actually blow through. Testing without it (A) verifies a different system. Verification is at least daily, not weekly (B). Filters are standard practice, not forbidden (C). Source: 2019 ATS/ERS spirometry standard, Calibration verification.
Q5. Plethysmograph accuracy validation
During monthly accuracy validation of an adult plethysmograph, an isothermal lung analog with a known volume of 4.00 L has a mean measured volume of 4.09 L across five determinations. How should this result be judged?
Show answer
Answer: C. For adult plethysmographs, the 2023 standard applies the ±50 mL or 3% limit, whichever is greater, to the mean of five determinations. For 4.00 L, 3% is 120 mL, so a 90-mL error passes. Option A applies the 50-mL figure without the "whichever is greater" rule. There's no 1% rule (B). Biological controls are a separate check (D). Source: 2023 ERS/ATS lung volumes standard, Body plethysmography: Quality control (PDF p. 9).
Q6. Helium dilution troubleshooting
Midway through a helium-dilution FRC, before equilibration, the helium concentration drops suddenly. What is the most likely problem, and what should be done?
Show answer
Answer: B. In the 2023 standard, a sudden drop in helium concentration signals a leak. Because the calculation assumes all the lost helium was diluted into the lungs, a leak overestimates FRC. Discard the trial. Repeat after at least twice the previous dilution time, or longer in severe obstruction. Slow mixing produces a gradual curve, not a sudden step (A). Equilibration means less than 0.02% change over 30 seconds, not an abrupt fall (D). Option C names a real maintenance issue but doesn't explain a sudden helium drop. Source: 2023 ERS/ATS lung volumes standard, Helium dilution: Measurement technique, steps 5–7 and Quality control (PDF p. 15); Acceptability and grading (PDF p. 17).
Q7. Nitrogen washout troubleshooting
During a multiple-breath nitrogen washout, end-tidal N2 has fallen to about 6% when one breath suddenly shows expired N2 near 30%. The patient's tidal breathing remains stable, but the technologist sees the mouthpiece seal break. What should the technologist conclude?
Show answer
Answer: C. The 2023 standard lists a sudden large rise in expired tracer concentration as evidence of a leak, and any evidence of leak means the trial is rejected. Room air is about 78% N2, so the observed break in the seal explains the abrupt rise. No cough, sigh or other breathing change is described to support release of trapped gas (A). The test ends when end-tidal N2 falls below 1/40 of its starting value for three consecutive breaths (B). Before repeating, wait at least twice the washout time. Applying an unexplained correction (D) does not repair a leak. Source: 2023 ERS/ATS lung volumes standard, Multiple-breath washout: Measurement technique, steps 5–8 (PDF pp. 11–12); Acceptability and grading (PDF p. 17).
Q8. Zero-flow problem
With nothing connected and no one breathing, a flow-sensor spirometer's display shows volume slowly accumulating. What is the correct fix?
Show answer
Answer: D. The 2019 update requires setting zero flow with the spirometer blocked before calibration checks and patient tests, and repeating it if the zero level shifts. A drifting baseline is a "faulty zero-flow setting," which makes maneuvers unacceptable and unusable. You can't assume the software handles it (A). BTPS is a temperature and humidity conversion, not a zero (B). The syringe isn't involved when no air is moving (C). Source: 2019 ATS/ERS spirometry standard, Zero-flow; Table 7.
Q9. Room temperature shift
The lab uses an older DLCO system without automatic monitoring or compensation for ambient changes. The lab's air conditioning fails, and room temperature rises from 21 °C to 25 °C within 20 minutes during a DLCO session. What should happen before the next patient?
Show answer
Answer: A. The 2017 DLCO standard says older systems may require another volume calibration check when room temperature changes by more than 3 °C. Here it changed 4 °C. The 2019 spirometry update notes that continuous temperature correction may be necessary when ambient temperature changes by more than 3 °C in under 30 minutes. BTPS conversion only works if the ambient temperature entered is correct (B). Stopping for the day (C) isn't required if the check passes. Barometric pressure isn't the variable that changed (D). Sources: 2017 ERS/ATS DLCO standard, Equipment calibration and quality control, item 2 (PDF p. 7); 2019 ATS/ERS spirometry standard, BTPS Adjustment.
Q10. Airway resistance and lung volumes
To shorten a full PFT, a technologist plans to record airway resistance and FRC from one panting maneuver in the plethysmograph. What is the best response?
Show answer
Answer: C. The 2023 lung volume standard states that airway resistance shouldn't come from the same panting maneuver used for lung volumes. The two measurements have different optimal panting frequencies, and staying on the mouthpiece longer increases leak risk. Option B is still one maneuver. Helium dilution can't measure airway resistance at all (D), and it changes the test. Source: 2023 ERS/ATS lung volumes standard, Plethysmography.
Q11. Inspiratory volume failure
A flow-sensor spirometer's daily check gives expiratory strokes of 2.98, 3.00 and 3.02 L and inspiratory strokes of 3.14, 3.16 and 3.17 L, using a 3.00-L syringe. What is the correct interpretation?
Show answer
Answer: B. The 2019 update requires the ±3% window for inspiration and expiration on flow-based devices; expiration-only applies to volume-based spirometers. Every inspiratory stroke is above 3.09 L. Averaging good and bad strokes (C) hides the failure. The expiratory strokes are fine (D). Inspiratory accuracy matters on flow sensors because maneuvers such as the FIVC check depend on it. Source: 2019 ATS/ERS spirometry standard, Calibration verification.
Q12. DLCO biological control
A lab's weekly DLCO biological control has a mean of 31.0 mL/min/mmHg from previous tests. Today it measures 26.8. What should happen?
Show answer
Answer: D. Today's value is 4.2 units (13.5%) below the mean. The 2017 standard says to repeat the test when a biological control deviates by more than 12%, or 3 mL/min/mmHg, from its previous mean. A confirmed deviation means the instrument needs evaluation before patients are tested. This exceeds the guideline's screening limits (A). Changing predicted values (B) or forcing a span (C) would conceal the discrepancy instead of finding its cause. Source: 2017 ERS/ATS DLCO standard, Equipment calibration and quality control, item 3 (PDF p. 8).
Q13. Metabolic cart before exercise testing
A metabolic cart was calibrated at 8 a.m. Before a cardiopulmonary exercise test (CPET) at 1 p.m., what calibration does the 2019 ERS CPET statement recommend?
Show answer
Answer: B. Flow sensors and gas analyzers both drift, so the ERS statement recommends calibrating them immediately before each test, not once a day (A). The flow sensor is checked with a certified 3-L syringe at three or more flow ranges and should agree within ±3%. Each gas sensor needs two calibration points, and the statement prefers two reference gases of known concentration over one gas plus room air (D). Warm-up is part of preparation, not a calibration (C). Source: Radtke T et al. ERS statement on standardisation of CPET in chronic lung diseases, 2019, Metabolic carts.
Q14. Blood gas analyzer quality control
The blood gas analyzer's scheduled quality-control run comes back outside its acceptability limits for PCO2. An arterial sample is waiting to be run. What should happen?
Show answer
Answer: D. The AARC blood gas guideline says out-of-limit QC should trigger equipment troubleshooting, and QC must be verified "in control" before specimens are analyzed, with corrective actions documented. Repeating or averaging (A) doesn't fix a biased instrument; it reproduces the bias. A comment on the report (B) doesn't make an out-of-control result usable. Running the sample before QC is fixed (C) gets the order backward. Source: Davis MD et al. AARC Clinical Practice Guideline: Blood gas analysis and hemoximetry, 2013, §8.3.1.3 (journal p. 1698; PDF p. 5).
Q15. Syringe DLCO test
For the weekly check, a technologist runs a DLCO maneuver using a 3-L syringe as the "patient." The system reports a DLCO of 1.4 mL/min/mmHg. What does this mean?
Show answer
Answer: B. A syringe has no alveolar–capillary membrane, so no CO should "disappear." The 2017 standard expects the absolute value of calibration-syringe DLCO to be below 0.5 mL/min/mmHg. A higher value means the system thinks CO was taken up, which points to a leak or an analyzer problem. The 3-unit figure (C) is the biological control limit, a different test. Hemoglobin adjusts predicted values for patients (D), not an equipment check. Source: 2017 ERS/ATS DLCO standard, Equipment calibration and quality control, item 3 (PDF p. 8); Table 1 (PDF p. 5).
Q16. Disposable flow sensors
A spirometer uses disposable flow sensors, one per patient. How should the daily calibration check be done?
Show answer
Answer: D. The 2019 update says that when disposable sensors are used, a new sensor from the patient-test supply should be tested each day. That way the check reflects the sensors patients actually use. A permanent reference sensor (A) checks something patients never touch. Factory calibration doesn't remove the daily requirement (B), and monthly checks are too infrequent (C). Source: 2019 ATS/ERS spirometry standard, Calibration verification.
Part 2 — Procedures (Q17–38)
Q17. Bronchodilator withholding (pediatric)
A 13-year-old arrives for pre- and post-bronchodilator spirometry. His parent says he used his albuterol inhaler 2 hours ago for mild wheeze. He is now stable, and the ordering clinician wants testing off bronchodilators and considers withholding clinically appropriate. What is the best action?
Show answer
Answer: B. The 2019 update lists a 4–6 hour hold for short-acting beta-agonists before bronchodilator responsiveness testing. Testing now risks a false-negative response, because the baseline may already be bronchodilated. Withholding depends on the test purpose and the clinician's instructions. If the ordering clinician needs the test done today, document the dose time on the report. Albuterol lasts hours, not one hour (A). Doubling the dose (C) changes the protocol, not the problem. Adolescents routinely perform this test (D). Source: 2019 ATS/ERS spirometry standard, Bronchodilator Responsiveness Testing and Table 8.
Q18. Methacholine contraindication
An adult is referred for a methacholine challenge. Baseline FEV1 is 1.38 L, which is 64% of predicted. What should the technologist do?
Show answer
Answer: C. The 2017 ERS methacholine standard lists baseline airflow limitation as a relative contraindication and calls for discussion with the ordering physician or laboratory medical director before proceeding. It is defined as FEV1 below 60% of predicted or below 1.5 L in adults, and this adult fails the absolute-volume test. A low starting dose (B) doesn't remove a contraindication. Deep breaths to TLC (D) are discouraged, because they blunt the response. Source: 2017 ERS methacholine standard, p. 4, Contraindications—Low FEV1 and Table 1; p. 2, Inhalation protocol. Coauthor-uploaded full text.
Q19. Choosing a lung volume method
A patient with severe emphysema and known large bullae is referred for lung volumes. Which method gives the most complete measurement of thoracic gas volume?
Show answer
Answer: D. Gas dilution and washout methods only measure gas that communicates with the airways. The 2023 standard lists underestimation of FRC in obstruction, bullae and emphysema as their known limitation (A, B). Plethysmography includes trapped and poorly ventilated gas. Spirometry can't measure TLC (C). One plethysmography caveat: keep panting near 0.5–1.0 Hz, because fast panting overestimates FRC in obstruction. Source: 2023 ERS/ATS lung volumes standard, pp. 4–5, Measurement of FRC and Table 1; pp. 7–8, body plethysmography.
Q20. Test order: DLCO and nitrogen washout
A full PFT includes spirometry, a single-breath DLCO and lung volumes by multiple-breath nitrogen washout. How should DLCO and the washout be ordered?
Show answer
Answer: C. Nitrogen washout has the patient breathe 100% oxygen. The 2023 lung volume standard recommends doing DLCO before any multiple-breath nitrogen washout, because leftover oxygen can lead to an underestimated DLCO. The same logic is why the 2017 DLCO standard asks patients to be off supplemental oxygen for at least 10 minutes before testing when that's clinically acceptable. Neither one minute (D) nor "right away" (A) follows the recommended test order or establishes adequate recovery from oxygen exposure. The two tests aren't independent (B). A sequence that fits the standard is spirometry, DLCO, nitrogen washout, bronchodilator if ordered, then repeat spirometry. Sources: 2023 ERS/ATS lung volumes standard, pp. 6–7, Sequence of lung function measurements; 2017 ERS/ATS DLCO standard, p. 9, Patient condition.
Q21. Six-minute walk baseline
An adult with chronic respiratory disease starting pulmonary rehabilitation needs a baseline six-minute walk distance that will be compared with a post-program test. What protocol fits the 2014 ERS/ATS standard?
Show answer
Answer: A. When you're measuring change, the standard recommends two 6MWTs and recording the best. Rest at least 30 minutes between them, until heart rate and SpO2 return to baseline. In COPD, the second walk averaged about 26 m further, a learning effect that would otherwise look like improvement after rehabilitation. Averaging (C) or using the first walk (D) keeps that bias. Skipping the second test (B) may be reasonable for one-off staging or after recent testing. Source: 2014 ERS/ATS field walking test standard, journal pp. 1430–1431, adult scope and Reliability and learning effect for 6MWD; p. 1439, Test repetition (PDF pp. 3–4 and 12).
Q22. Exercise challenge setup (pediatric)
A 12-year-old soccer player gets chest tightness after games, and an exercise challenge is ordered. Which setup best follows the 2018 ERS indirect challenge standard?
Show answer
Answer: C. Exercise challenge is an indirect test. Breathing large volumes of dry air dries the airway surface, which triggers mediator release and bronchoconstriction. The protocol therefore uses dry air, a nose clip (so all air passes the mouth), a ramp of 2–3 minutes, and sustained high intensity for at least 4 minutes, preferably 6. The preferred target is 60% of maximum voluntary ventilation; a heart rate above 85% of predicted maximum can serve as a surrogate. FEV1 is then measured repeatedly over the next 30 minutes. Humid air at low intensity (A) weakens the stimulus. Three minutes without a nose clip (B) is too short and lets air be conditioned in the nose. Option D isn't the ordered test. Sources: 2018 ERS indirect airway challenge standard, pp. 6–7, Exercise challenge and Table 2; p. 8, Post-challenge spirometry protocol. The dry-air value is specified in the 2013 ATS exercise-induced bronchoconstriction guideline, journal p. 1019 (PDF p. 4), Exercise Challenge Testing to Identify EIB.
Q23. Back-extrapolated volume
On a forced maneuver, FVC is 3.00 L and the back-extrapolated volume (BEV) is 0.18 L. How should the maneuver be judged?
Show answer
Answer: D. The 2019 rule is BEV no more than 5% of FVC or 0.100 L, whichever is greater. Here that's 0.15 L, so 0.18 L fails. A high BEV means a hesitant, slow start, which makes FEV1 unreliable. That affects both FEV1 and FVC acceptability and usability. There's no 0.20-L limit (A). The 0.100-L floor applies at all ages (B). Repeatability doesn't rescue an unacceptable start (C). Source: 2019 ATS/ERS spirometry standard, Within-Maneuver Evaluation, Table 7 and the BEV discussion.
Q24. Cough during the maneuver
A patient coughs at 0.6 seconds into an otherwise good forced exhalation, which meets the end-of-forced-expiration criteria. How do the 2019 criteria treat this maneuver?
Show answer
Answer: A. In Table 7 of the 2019 update, "no cough in the first second" is an FEV1-only criterion, required for both acceptability and usability. A cough inside the first second disqualifies FEV1 entirely. FVC has its own criteria (BEV, zero flow, no glottic closure, end-of-forced-expiration and others), so the same maneuver can still contribute an acceptable FVC. Option B overextends the rule. Options C and D underapply it. Source: 2019 ATS/ERS spirometry standard, Table 7 and Within-Maneuver Evaluation, first-second cough.
Q25. Ending the exhalation
An adult exhales for 9 seconds. In the final second, the volume rises by 0.020 L. What is the end-of-forced-expiration (EOFE) status?
Show answer
Answer: B. The 2019 update defines EOFE by any of three indicators. Table 7 specifies a plateau of no more than 0.025 L of volume change in the last second. Reaching 15 seconds of forced expiratory time also counts. The third indicator, for patients who can't exhale long enough, is an FVC within the repeatability tolerance of or greater than the largest prior FVC. There is no minimum exhalation time (A, C). Fifteen seconds is a point at which EOFE is reached, not a requirement. Additional 15-second efforts are not required to establish a plateau already demonstrated (D). The older 6-second minimum often seen in textbooks no longer applies. Source: 2019 ATS/ERS spirometry standard, Table 7 (EOFE indicators) and Within-Maneuver Evaluation.
Q26. Panting frequency
A patient with moderate airflow obstruction pants at about 2.2 Hz in the plethysmograph. What should the technologist do?
Show answer
Answer: C. The 2023 standard targets 0.5–1.0 Hz. It rejects trials above 2.0 Hz, and above 1.5 Hz when spirometry shows significant obstruction. At fast frequencies, mouth pressure underestimates the change in alveolar pressure in obstructed lungs, so FRC is overestimated. Good repeatability (B) can't fix a systematic error. Panting below 0.5 Hz (D) causes problems with the box's controlled leak. Source: 2023 ERS/ATS lung volumes standard, pp. 7–8, Measurement of FRC using body plethysmography; Table 2, p. 17.
Q27. Linked maneuvers after FRC
Immediately after an acceptable plethysmographic FRC, what maneuver sequence does the 2023 standard recommend?
Show answer
Answer: B. The FRC is linked to an IC (giving TLC) and then a slow expiratory VC (giving RV). Forced expirations (A) can cause dynamic airway collapse and air trapping in obstructed patients, which overestimates RV. The 2023 standard doesn't recommend unlinked spirometry for plethysmography (C). TLC must be measured, not estimated (D). Source: 2023 ERS/ATS lung volumes standard, pp. 4–5, Derivation of lung subdivisions (linked FRC, IC and slow EVC); Figure 3, p. 6.
Q28. DLCO inspired volume
In a single-breath DLCO using a classical (discrete-sample) system, the patient's inspired volume (VI) is 3.40 L. Her largest vital capacity this session is 4.20 L. Breath-hold time is 10 seconds. What is the best action?
Show answer
Answer: D. VI is 3.40 ÷ 4.20 = 81% of VC. The 2017 standard wants at least 90%, or at least 85% with VA within 200 mL or 5% (whichever is greater) of the largest VA from another acceptable maneuver. This 81% inspired volume fails both acceptability routes; reduced inspiration can affect the measurement. An 81% VI/VC is not grade A (A), and VI clearly matters (B). The test gas must wash out before the next maneuver: wait at least 4 minutes on classical systems, and consider 10 minutes in airflow obstruction (C). Source: 2017 ERS/ATS DLCO standard, pp. 9–10, Inspiratory manoeuvres; p. 12, Manoeuvre intervals in classical systems; Table 3 and footnote, p. 22 (corrected November 2018 version).
Q29. Methacholine inhalation technique
Under the 2017 ERS standard, how should a patient inhale each methacholine dose using a continuous or breath-actuated nebulizer?
Show answer
Answer: B. Deep inspirations to TLC have a bronchoprotective, bronchodilating effect that blunts the response. The 2017 standard therefore favors at least one minute of tidal breathing with these nebulizers; dosimeters instead use a calculated breath count. It reports the delivered dose (PD20). Options A and D build deep breaths into every dose. Option C changes deposition and isn't a recognized method. Source: 2017 ERS methacholine standard, p. 2, Inhalation protocol; pp. 9–10, Duration of nebulisation and Procedure steps 4–5. Coauthor-uploaded full text.
Q30. Patient stops during a 6MWT
Three minutes and 40 seconds into a six-minute walk, an adult with chronic respiratory disease stops and leans on the wall with tolerable breathlessness. SpO2 is 89% with a quality signal, and there's no chest pain or other stopping criterion. What should the assessor do?
Show answer
Answer: C. In the 2014 standard, the clock never stops. Patients may rest standing or sitting and should resume when able. Encouragement is limited to the standard phrases, every 30 seconds once SpO2 is 85% or higher. Record the time stopped and the number of stops. The assessor stops the test for SpO2 below 80% or concerning symptoms, including intolerable dyspnea; no stopping criterion is present here (B). Pausing the timer (A) invalidates the distance. Setting the pace (D) is explicitly avoided; assessors walk behind. Source: 2014 ERS/ATS field walking test standard, journal p. 1439, Reasons for the assessor to stop an exercise test; p. 1440, 6MWT Conduct and Table 6 (PDF pp. 12–13).
Q31. Before radial puncture
Before a radial arterial puncture, a modified Allen test is performed. After the clinician releases the ulnar artery, the palm remains pale at 15 seconds. What should happen?
Show answer
Answer: D. A palm that doesn't regain color suggests inadequate collateral flow through the ulnar artery. Puncturing that radial artery is then considered unsafe, because injury could compromise blood supply to the hand. Pressure or needle size (A, B) doesn't change the collateral-flow problem. Repeating the test until it "passes" (C) ignores the finding. Source: WHO guidelines on drawing blood: best practices in phlebotomy (2010), Annex I, Modified Allen test, printed p. 85 (PDF p. 101); section 5.2.2, printed p. 32.
Q32. Repeatability after three good maneuvers
An adult has three acceptable maneuvers. The two largest FVC values are 4.12 and 3.95 L. The two largest FEV1 values are 3.20 and 3.12 L. What should the technologist do?
Show answer
Answer: A. For patients older than 6, the 2019 update requires the two largest FVCs and the two largest FEV1s to agree within 0.150 L. FEV1 passes (0.08 L), but FVC doesn't (0.17 L). Keep trying, up to about eight maneuvers for adults, and stop sooner if the patient can't continue or FEV1 from an acceptable maneuver falls below 80% of its starting value. FEV1 alone isn't enough (B). The standard reports the largest acceptable values, not averages (C). Repeatability is assessed as the maneuvers are obtained, not delayed until fatigue (D). Source: 2019 ATS/ERS spirometry standard, Between-Maneuver Evaluation, Maximum Number of Maneuvers and Reported Values.
Q33. Forced inspiratory check
A maneuver shows FVC 3.40 L and a forced inspiratory vital capacity (FIVC), measured right after, of 3.62 L. How should it be judged?
Show answer
Answer: C. The 2019 limit is FIVC − FVC ≤ 0.100 L or 5% of FVC, whichever is greater. Here 5% of FVC is 0.17 L, and the difference is 0.22 L. The maneuver didn't start from full inspiration, so FVC and FEV1 are understated. Coach a complete inhalation before the blast. The 0.25-L figure (A) doesn't exist. The check applies to everyone (B). Option D reverses the logic: FIVC can equal or slightly exceed FVC, just not by more than the limit. Source: 2019 ATS/ERS spirometry standard, Maximal inspiration after forced expiration; Within-Maneuver Evaluation and Table 7 (FIVC minus FVC).
Q34. DLCO repeatability
On a classical (discrete-sample) system, two acceptable single-breath DLCO maneuvers give 24.1 and 27.0 mL/min/mmHg. The patient is comfortable and able to continue. What is the best next step?
Show answer
Answer: B. The 2017 standard requires at least two acceptable maneuvers within 2 mL/min/mmHg. These differ by 2.9. Once repeatable maneuvers are obtained, the mean is reported. More than five maneuvers are not recommended because each test raises COHb (about 3.5% over five) and lowers measured DLCO. After only two attempts, seek better repeatability before reporting the mean (A). If repeatability cannot be achieved after further testing, report the average acceptable DLCO with a quality caution. Reporting the higher value (C) is not the reporting rule. Retesting without the washout interval (D) carries over test gas. Source: 2017 ERS/ATS DLCO standard, p. 12, Manoeuvre intervals in classical systems; pp. 21–22, Repeatability, suboptimal-session reporting and Table 3 (corrected November 2018 version).
Q35. Plethysmographic FRC repeatability
Three acceptable FRCpleth values are 3.10, 3.22 and 3.30 L. Which statement is correct?
Show answer
Answer: D. The 2023 repeatability test is (highest − lowest) ÷ mean. That gives 0.20 ÷ 3.207 = 6.2%, above the 5% target for three values. Keep testing for three that agree within 5%; the standard notes that values with poorer repeatability may still be useful if fatigue ends the session. Checking pairs (B) isn't the criterion. Reported FRC is a mean of acceptable values, not the highest (C). Source: 2023 ERS/ATS lung volumes standard, Repeatability (PDF p. 16); Reporting—Selection of manoeuvres (PDF pp. 18–19).
Q36. Pulse oximetry after smoke exposure
A patient pulled from a house fire has a conventional two-wavelength pulse-oximeter SpO2 of 98%. Why is a blood CO-oximetry (hemoximetry) measurement needed?
Show answer
Answer: B. Dyshemoglobins such as carboxyhemoglobin change the light absorption that conventional two-wavelength oximeters rely on, which produces false readings. A CO-oximeter measures the hemoglobin species directly. The error runs toward overestimation, not underestimation (A). Option C ignores COHb. CO-oximetry measures hemoglobin fractions, not PaO2 (D). Sources: CDC clinical guidance for carbon monoxide poisoning, Confirmation of Diagnosis; Buckley et al., The pulse oximetry gap in carbon monoxide intoxication (1994), Abstract—Results and Conclusion.
Q37. Preschool repeatability (pediatric)
A 5-year-old's two largest acceptable FEV1 values are 1.02 L and 0.94 L. How should FEV1 repeatability be judged?
Show answer
Answer: C. For children 6 and younger, the 2019 update allows 0.100 L or 10% of the highest value, whichever is greater. Ten percent of 1.02 L is 0.102 L, so the limit is 0.102 L, and the 0.08-L difference passes. The 0.150-L rule (B) is for patients older than 6. FEV1 and FVC repeatability are judged separately. There's no 5% rule (A) or exact-match rule (D). Source: 2019 ATS/ERS spirometry standard, Table 7; Between-Maneuver Evaluation.
Q38. Air in the blood gas syringe
An arterial sample from a patient whose true PaO2 is about 55 mm Hg is kept at room temperature for 5 minutes before analysis. A large room-air bubble was never expelled from the syringe. What error pattern does the bubble most likely cause?
Show answer
Answer: A. Room air has a higher PO2 than this patient's arterial blood and almost no CO2. Gas equilibrates across the bubble, so PO2 drifts up and PCO2 drifts down. Patterns B and C run the wrong direction for air contamination. Prompt processing does not make an air-contaminated specimen acceptable (D); removing the bubble later does not restore the original gas values. The AARC blood gas guideline lists a specimen with visible air bubbles as one not to analyze, so a sample like this should be flagged under your lab's specimen-rejection protocol. Sources: Biswas et al., Blood gas analysis: effect of air bubbles in syringe and delay in estimation (1982), Abstract—air-bubble experiment results; AARC blood gas and hemoximetry guideline, 2013, §4.4 (printed p. 1695; PDF p. 2) and §6.1.2.1 (printed p. 1696; PDF p. 3).
Part 3 — Data management (Q39–50)
Q39. Bronchodilator response
Pre-bronchodilator FEV1 is 1.20 L and post-bronchodilator FEV1 is 1.42 L. Predicted FEV1 is 3.00 L. How does the 2022 ERS/ATS interpretation standard classify the FEV1 response?
Show answer
Answer: B. The 2022 standard defines a significant bronchodilator response as a change of more than 10% of the predicted value for FEV1 or FVC: 0.22 ÷ 3.00 = 7.3%. The older "12% and 200 mL from baseline" rule (A) would call this positive. This is a case where the two methods disagree. No 200-mL-only rule exists (C), and the 2022 criterion needs only the predicted value, not a z-score for the change (D). Source: 2022 ERS/ATS interpretation standard, Bronchodilator responsiveness (PDF pp. 10–11); Box 1 (PDF p. 11).
Q40. Hemoglobin-adjusted predicted DLCO
An adult man has a predicted DLCO of 30.0 mL/min/mmHg and a hemoglobin of 10.0 g/dL. Using the 2017 ERS/ATS formula, what is his hemoglobin-adjusted predicted DLCO?
Show answer
Answer: A. For adolescent and adult males: predicted DLCO × [1.7 × Hb ÷ (10.22 + Hb)] = 30.0 × (17 ÷ 20.22) = 30.0 × 0.841 = 25.2. Anemia lowers the predicted value. Option C uses the constant for females and children under 15 (9.38). Option B ignores the anemia. Option D divides instead of multiplies. The 2017 standard applies this adjustment to the predicted value, not the measured one. Source: 2017 ERS/ATS DLCO standard, Adjustment for haemoglobin; equations 31–32 (PDF p. 23).
Q41. Severity by z-score (pediatric)
A 16-year-old's technically acceptable spirometry shows FEV1/FVC of 0.74, with a z-score of −2.1, and an FEV1 z-score of −3.2. How should the physiological impairment be classified under the 2022 standard?
Show answer
Answer: C. Obstruction is FEV1/FVC below the lower limit of normal (z below −1.645), and −2.1 qualifies even though the ratio is 0.74. The fixed 0.70 cutoff (A) is strongly discouraged, because it misclassifies young and old patients. Severity is graded by the FEV1 z-score. The moderate range runs from −2.51 to −4. At −3.2, this is moderate physiological impairment. This grade describes lung-function impairment, not overall disease severity. Restriction (D) can't be called from spirometry; it needs a TLC below the LLN. Source: 2022 ERS/ATS interpretation standard, Severity of lung function impairment (PDF pp. 13–15); Figure 7 (PDF p. 14); Table 5 (PDF p. 18); Restrictive impairments (PDF p. 21).
Q42. Calculating RV
For the same selected acceptable and repeatable plethysmographic maneuvers, mean FRC is 3.40 L and mean linked inspiratory capacity is 2.10 L. The linked slow VC is 3.50 L; the largest VC from any acceptable maneuver this session is 3.60 L. What RV should be reported?
Show answer
Answer: A. Mean TLC = mean FRC + mean linked IC for the same selected maneuvers = 3.40 + 2.10 = 5.50 L (D is the TLC, not the RV). The 2023 standard reports RV as TLC minus the largest VC: 5.50 − 3.60 = 1.90 L. Option C uses the smaller linked VC (5.50 − 3.50), which overstates RV. Option B subtracts IC from FRC, which isn't a lung volume. Source: 2023 ERS/ATS lung volumes standard, Derivation of lung subdivisions (PDF p. 5); Table 3 and Reporting—Selection of manoeuvres (PDF p. 18).
Q43. Diluent step
Baseline FEV1 is 3.00 L. After the diluent (saline) step of a methacholine challenge, FEV1 is 2.34 L. Both values come from technically acceptable maneuvers. What is the correct action?
Show answer
Answer: D. The fall is 0.66 ÷ 3.00 = 22%. A fall of 20% or more after diluent meets the standard's cancellation criterion, with any needed rescue bronchodilator and recovery checks. Continuing (A, B) or escalating (C) would ignore that stopping rule. Source: 2017 ERS methacholine standard, p. 10, Procedure steps 6–7; p. 5, Precautions for patient safety; p. 13, Data presentation. Coauthor-uploaded full text.
Q44. VA versus TLC
An adult patient's single-breath DLCO is reduced. Both the DLCO maneuver and same-day plethysmographic lung-volume measurements meet quality criteria. VA is 3.9 L, and TLC is 5.8 L. How should this affect the interpretation?
Show answer
Answer: D. VA ÷ TLC = 3.9 ÷ 5.8 = 0.67. The 2022 standard says a ratio well below about 0.85–0.90 suggests gas maldistribution — the test gas didn't reach all of the lung — so DLCO may be underestimated. A healthy VA is close to TLC, so A is wrong. Swapping volumes (B) isn't a recognized correction. KCO rises as lung volume falls, so a "normal" KCO doesn't cancel a low VA (C). Source: 2022 ERS/ATS interpretation standard, Gas transfer impairments: VA/TLC and interpretation of KCO at low VA (PDF p. 23).
Q45. End of exhalation in a child (pediatric)
A 9-year-old can't exhale longer than 2.8 seconds, despite coaching, and shows no plateau. This maneuver's FVC is 2.05 L, and the largest FVC from her earlier maneuvers in the same testing set is 2.12 L. All other criteria are met. Is the FVC acceptable?
Show answer
Answer: A. The 2019 update's third EOFE indicator is for patients who can't exhale long enough to plateau, such as children with high elastic recoil. Their FVC counts if it is within repeatability tolerance of, or greater than, the largest prior FVC in the same testing set. For a 9-year-old the tolerance is 0.150 L, and 2.12 − 2.05 = 0.07 L. The plateau isn't the only route (B), and there's no minimum time (C). Because an EOFE acceptability criterion is met, the FVC can be acceptable rather than merely usable (D). Source: 2019 ATS/ERS spirometry standard, Table 7, third EOFE criterion and footnote; Within-Maneuver Evaluation.
Q46. Smoking before DLCO
An outpatient tells you he smoked a cigarette 1 hour before his DLCO. How does this affect the result, and what should be done?
Show answer
Answer: C. The 2017 standard asks patients not to smoke on test day. It requires recording the time of the last cigarette and correcting for CO back-pressure after recent or heavy smoking, because elevated COHb lowers measured DLCO. That reverses B and contradicts A. Oxygen right before the test (D) also biases DLCO downward, and the standard asks for no supplemental O2 for 10 minutes beforehand when that's clinically acceptable. Source: 2017 ERS/ATS DLCO standard, Patient condition (PDF p. 9); Adjustments for carboxyhaemoglobin concentration and carbon monoxide back-pressure (PDF pp. 24–25).
Q47. Reduced FVC with normal TLC
Technically acceptable spirometry shows FVC z −2.3, FEV1 z −2.0 and FEV1/FVC z +0.2. Plethysmographic TLC has a z-score of −0.6. What is the best description?
Show answer
Answer: B. A reduced FVC and FEV1 with a normal ratio and a normal TLC is what the 2022 standard calls a non-specific pattern. It can reflect submaximal effort, early small-airway disease or other causes. Restriction (A) requires TLC below the LLN. The ratio is normal, so the measured pattern is neither spirometric obstruction (C) nor a mixed ventilatory impairment (D). Source: 2022 ERS/ATS interpretation standard, Table 5 (PDF p. 18); Non-specific pattern (PDF p. 19).
Q48. Low ratio and low TLC
Technically acceptable results show FEV1/FVC z −2.4 and plethysmographic TLC z −2.0. What is the best description?
Show answer
Answer: D. Both the ratio and TLC are below the LLN (z under −1.645). The 2022 standard defines that combination as a mixed defect, and it can be confirmed only with a measured TLC. Option A misses the restrictive component established by the low TLC. The ratio being below the LLN isn't dismissed (B). A non-specific pattern (C) requires a normal ratio and a normal TLC. Source: 2022 ERS/ATS interpretation standard, Mixed ventilatory impairments and Table 7 (PDF p. 22).
Q49. Upright versus supine vital capacity
A patient's acceptable and repeatable slow VC measurements are 3.60 L sitting and 2.30 L supine. What does this most suggest?
Show answer
Answer: A. Lying down shifts abdominal contents against the diaphragm. A healthy diaphragm handles this with only a small loss of VC, but a weak one can't, so VC falls markedly. This drop is about 36%, far beyond a normal postural change (D). The ATS/ERS respiratory-muscle statement uses the upright-to-supine VC fall as a sign of diaphragm weakness. Obstruction (B) isn't assessed by this comparison. VC does change with position (C), which is the whole point of the test. This finding warrants further assessment; it does not establish the diagnosis by itself. Sources: ATS/ERS Statement on Respiratory Muscle Testing (2002), Static Lung Volumes (printed p. 521; PDF p. 4); Fromageot et al., Supine fall in lung volumes in the assessment of diaphragmatic weakness in neuromuscular disorders (2001), Abstract—Results and Conclusions; study of 24 patients with neuromuscular disease.
Q50. Methacholine result category
A methacholine challenge reports a PD20 of 60 µg. How is this categorized under the 2017 ERS standard?
Show answer
Answer: D. The 2017 categories by delivered dose are: over 400 µg normal; 100–400 µg borderline; 25–100 µg mild; 6–25 µg moderate; under 6 µg marked. At 60 µg, the result is mild. The interpretation still depends on the clinical picture, symptoms and medications held. The standard reports PD20 (delivered dose) instead of PC20 (concentration), so results can be compared across nebulizers when their delivery characteristics and dosing protocols are known. Source: 2017 ERS methacholine standard, Table 6, p. 14; pp. 13–14, Interpretation of direct bronchial challenge tests; p. 2, Delivery devices. Coauthor-uploaded full text.
Score your practice test
Count your correct answers in each part. An unanswered question counts as not correct: 30 right with 5 skipped is 30/50, not 30/45. The three parts follow the domains of the PFT content outline, in approximately the same proportions as the 100 scored items on the real exam.
| Domain | Questions here | Scored items on the real exam | Your correct |
|---|---|---|---|
| Instrumentation / equipment | Q1–16 (16) | 33 | ___ / 16 |
| Procedures | Q17–38 (22) | 44 | ___ / 22 |
| Data management | Q39–50 (12) | 23 | ___ / 12 |
| Total | 50 | 100 | ___ / 50 |
Your result describes these 50 questions only. It isn't an NBRC exam score, and it doesn't predict whether you'd reach the CPFT or RPFT cut score. With 12–22 questions per domain, a part score points you toward a weak area; it isn't a precise measurement.
Turn misses into a study list. Most wrong answers trace back to one document, so reread the rule, not just the question:
| If you missed… | Reread |
|---|---|
| Q1, 4, 8, 9, 11, 16, 17, 23–25, 32, 33, 37, 45 | 2019 ATS/ERS spirometry standard: calibration, acceptability, repeatability |
| Q3, 9, 12, 15, 20, 28, 34, 40, 46 | 2017 ERS/ATS DLCO standard |
| Q2, 5–7, 10, 19, 20, 26, 27, 35, 42 | 2023 ERS/ATS lung volume standard |
| Q39, 41, 44, 47, 48 | 2022 ERS/ATS interpretation standard |
| Q18, 29, 43, 50 | 2017 ERS methacholine standard: contraindications, dosing, diluent response, PD20 |
| Q22 | 2018 ERS indirect challenge standard: exercise challenge, Table 2 |
| Q13 | 2019 ERS cardiopulmonary exercise testing statement: equipment calibration, p. 14 |
| Q21, 30 | 2014 ERS/ATS field walking test standard: six-minute walk methodology |
| Q49 | ATS/ERS respiratory muscle statement: static lung volumes, p. 521 |
| Q14, 38 | AARC blood gas and hemoximetry guideline: analyzer QC and sample rejection |
| Q31 | WHO guidelines on drawing blood: Annex I, modified Allen test |
| Q36 | CDC carbon monoxide guidance: confirmation of diagnosis |
Then pick two missed questions. Explain out loud why the right answer fits and which rule you confused, and try them again tomorrow without looking at the explanation.
Want a timed run? The real exam allows 2 hours for 115 questions, about 1 minute 3 seconds each. At that pace, these 50 take about 52 minutes.
NBRC PFT exam at a glance
| Exam detail | Current NBRC information |
|---|---|
| Exam | Pulmonary Function Technology (PFT) Examination |
| Credentials | Low cut score → CPFT · High cut score → RPFT |
| Questions | 115 multiple-choice (100 scored, 15 unscored pretest) |
| Time | 2 hours |
| Content outline | Effective October 2022: Instrumentation/Equipment 33 · Procedures 44 · Data Management 23 scored items; 83 of the 100 test application or analysis rather than recall |
| Patient focus | 10 pediatric and 90 general items per form |
| Eligibility | Age 18+ and one of: an associate degree or higher from a CoARC-accredited or supported entry-into-practice respiratory therapy program; the NBRC CRT or RRT credential; 62 semester hours of college credit from a college or university accredited by its regional association or equivalent, including college-level biology and mathematics; or the CPFT credential if seeking RPFT |
| Fee | $200 new applicants · $170 repeat applicants (AARC members get a one-time $40 discount on a first PFT application, registered through the AARC) |
| Where | PSI assessment centers, Monday–Saturday except nationally recognized holidays; schedule online once eligibility is confirmed |
| Score | Pass/fail appears on screen after the exam; PSI emails the score report within 24 hours |
| Retakes | Two attempts without waiting; after that, at least 180 days between attempts |
Sources: NBRC Candidate Handbook, revised September 3, 2026, printed pp. 4–5, 8, 12–13; NBRC Resources FAQs, PFT eligibility, AARC discount and score delivery; and the PFT content outline.
Some third-party pages still list chemistry or six months of experience as requirements. The NBRC's current admission requirements include neither. Fees and policies can change, so check the NBRC page before you apply.
CPFT or RPFT: how the two cut scores work
There is one PFT exam with two cut scores. Reach the low cut and you earn the CPFT. Reach the high cut and you earn the RPFT. You don't take a separate RPFT test.
If you already hold the CPFT, you retake the same exam to reach the high cut. Scoring below the low cut doesn't cost you your CPFT.
The high cut is a real step up. In 2025, 68.8% of new candidates reached the low cut and 31.5% reached the high cut. Among repeat candidates, the figures were 54.2% and 15.7% (NBRC 2025 Examinations in Review, p. 2).
For retake timing and what your score report shows, see the NBRC retake policy and NBRC exam results.
The standards these questions are built on
The NBRC says its PFT exam committee relies on current ATS and ERS technical standards. It warns that popular textbooks may not keep pace, and it publishes the list candidates should use. These questions draw on the standards below, with additional sources cited beside individual answers:
- Spirometry: ATS/ERS Standardization of Spirometry, 2019 update
- DLCO: ERS/ATS single-breath CO uptake standards, 2017
- Lung volumes: ERS/ATS lung volume technical standard, 2023 update
- Interpretation: ERS/ATS interpretive strategies, 2022
- Bronchial challenge: ERS methacholine standard, 2017 and the ERS indirect challenge standard, 2018
- Field walking tests: ERS/ATS technical standard, 2014
- Respiratory muscle physiology: ATS/ERS statement, 2002, “Static Lung Volumes,” for the postural vital-capacity question
Rules that trip people up:
- [ ] Under the 2022 ERS/ATS interpretation standard, airflow obstruction is FEV1/FVC below the LLN (5th percentile), rather than a universal 0.70 cutoff (Q48).
- [ ] Lung function impairment severity is graded by z-score under the 2022 ERS/ATS interpretation standard (Q41).
- [ ] Under the 2022 ERS/ATS standard, a positive bronchodilator response is an increase in FEV1 or FVC of more than 10% of its predicted value, rather than the older “12% and 200 mL” criterion (Q39).
- [ ] There is no minimum forced expiratory time for FVC acceptability. A plateau of 0.025 L or less over at least 1 second meets Table 7's end-of-forced-expiration criterion. When a patient cannot achieve a plateau, FVC can satisfy that criterion by being greater than or within repeatability tolerance of the largest prior FVC in the same testing set; other acceptability criteria still apply (Q25, Q45).
- [ ] Nitrogen washout ends when end-tidal N2 is below 1/40 of its starting value for at least three consecutive tidal breaths (Q7).
- [ ] For plethysmography, aim for at least three acceptable FRC values with (highest − lowest) ÷ mean ≤5%. Link FRC to an inspiratory capacity and then a slow expiratory VC without leaving the mouthpiece. Lower-quality results may still be usable when the target cannot be achieved (Q27, Q35).
- [ ] Methacholine dosing follows the device-specific tidal-breathing or dosimeter protocol; deep inhalations to TLC during dosing are not recommended. Results use PD20, the delivered dose causing a 20% fall in FEV1 (Q29, Q50).
- [ ] For DLCO, inspired volume should be at least 90% of the largest VC in the same session; the ≥85% exception requires VA within 200 mL or 5%, whichever is greater, of the largest VA from another acceptable maneuver. The calculated breath-hold is 10 ±2 seconds, and repeatability requires at least two acceptable measurements within 2 mL/min/mmHg. More than five maneuvers is not recommended. When both tests are scheduled, perform DLCO before multiple-breath nitrogen washout (Q20, Q28, Q34).
Where to go next
- Take NBRC's free PFT practice exam. Use the official resource to familiarize yourself with the content and question types. Choose "PFT Practice Exam" on NBRC's free practice exams page.
- Use the content outline as a checklist. The PFT Detailed Content Outline maps the examination's tasks.
- Apply and schedule. Apply through the NBRC once you meet one of the four eligibility pathways, then book a PSI center at test-takers.psiexams.com/nbrc.
Sources
- NBRC, Candidate Handbook, revised September 3, 2026, printed pp. 4–5, 8, 12–13; Resources FAQs, PFT admission requirements, fees, discounts and score delivery.
- NBRC, PFT Detailed Content Outline, effective October 2022.
- NBRC, ATS and ERS Update Standards Critical for Exam Preparation.
- NBRC, Free Practice Exams.
- NBRC, 2025 Examinations in Review.
- Graham BL et al. Standardization of Spirometry 2019 Update. Am J Respir Crit Care Med 2019;200(8):e70–e88.
- Stanojevic S et al. ERS/ATS technical standard on interpretive strategies for routine lung function tests. Eur Respir J 2022;60:2101499.
- Graham BL et al. 2017 ERS/ATS standards for single-breath carbon monoxide uptake in the lung. Eur Respir J 2017;49:1600016.
- Bhakta NR et al. ERS/ATS technical standard: standardisation of the measurement of lung volumes, 2023 update. Eur Respir J 2023;62:2201519.
- Coates AL et al. ERS technical standard on bronchial challenge testing: general considerations and performance of methacholine challenge tests. Eur Respir J 2017;49:1601526. Author-posted full text.
- Holland AE et al. ERS/ATS technical standard: field walking tests in chronic respiratory disease. Eur Respir J 2014;44:1428–46.
- Hallstrand TS et al. ERS technical standard on bronchial challenge testing: pathophysiology and methodology of indirect airway challenge testing. Eur Respir J 2018;52:1801033, Table 2 and pp. 6–8.
- Parsons JP et al. ATS clinical practice guideline: exercise-induced bronchoconstriction. Am J Respir Crit Care Med 2013;187:1016–1027, p. 1019, dry-air condition.
- ATS/ERS, Statement on Respiratory Muscle Testing. Am J Respir Crit Care Med 2002;166:518–624, “Static Lung Volumes,” p. 521. Fromageot C et al., Supine fall in lung volumes in the assessment of diaphragmatic weakness in neuromuscular disorders. Arch Phys Med Rehabil 2001;82:123–128.
- Radtke T et al. ERS statement on standardisation of cardiopulmonary exercise testing in chronic lung diseases. Eur Respir Rev 2019;28:180101.
- Davis MD et al. AARC Clinical Practice Guideline: Blood gas analysis and hemoximetry: 2013. Respir Care 2013;58(10):1694–1703.
- WHO, Guidelines on drawing blood: best practices in phlebotomy, 2010, Annex I, p. 85.
- CDC, Clinical Guidance for Carbon Monoxide Poisoning, August 21, 2026, “Confirmation of Diagnosis.” Buckley RG et al., The pulse oximetry gap in carbon monoxide intoxication. Ann Emerg Med 1994;24:252–255.
- Biswas CK et al., Blood gas analysis: effect of air bubbles in syringe and delay in estimation. BMJ 1982;284:923–927.
Last verified: Exam facts were checked October 30, 2026 against the NBRC Candidate Handbook (revised September 3, 2026), NBRC Resources FAQs, the PFT Detailed Content Outline (effective October 2022) and the NBRC 2025 Examinations in Review. Each question's underlying rule was checked against the source cited beneath it.
By the Castleport Test Prep Editorial Team.
We use AI-assisted tools in our editorial workflow. These source checks are not professional clinical review. See our methodology.
Castleport Test Prep is an independent exam prep publisher. It is not affiliated with, endorsed by or approved by the National Board for Respiratory Care (NBRC). Exam and credential names identify their subjects, and trademarks belong to their respective owners. These practice questions are original and unofficial; they are not NBRC exam or practice items. Use of this page doesn't guarantee a passing score or any credential.