This free guide follows ARRT's Mammography content specifications in effect since September 1, 2025. The exam has 115 scored questions plus 30 unscored pilot questions, with 150 minutes of test time, and you pass with a scaled score of 75.

Study priority ARRT major category ARRT subcategory Scored questions Share of scored exam
1 Procedures Mammographic Positioning and Procedures 39 34%
2 Image Production Image Acquisition and Quality Assurance 30 26%
3 Procedures Anatomy, Physiology, and Pathology 26 23%
4 Patient Care Patient Interactions and Management 20 17%
Total Total 115 100%

Counts from ARRT's Mammography Examination Content Specifications, p. 1. Shares are our arithmetic (category ÷ 115, rounded). Positioning and anatomy together make up ARRT's Procedures category: 65 of 115 scored questions.

Start with Question 1 below, then use the study plan to cover every category.

Practice questions: 25 original items weighted like the exam

These are original, unofficial practice questions written by Castleport Test Prep. They are not ARRT exam questions. The set follows the scored-question shares above: 8 positioning and procedures, 6 anatomy and pathology, 7 image production, and 4 patient care. Your result is feedback on these 25 items, not a scaled score or a prediction.

The real exam is mostly single-best-answer multiple choice. ARRT also uses some select-multiple, sorted-list, select-an-image, and video-based items (2026 Postprimary Handbook, p. 39). Every item here is single best answer.

All 25 answers are open. Return to the normal view

Positioning and procedures (8 questions)

Question 1 · CC/MLO evaluation: posterior nipple line

On a patient's right MLO, the posterior nipple line (PNL) measures 11.0 cm. Which right CC measurement shows adequate posterior tissue on the CC?

  • A. 8.0 cm
  • B. 12.5 cm
  • C. 10.2 cm
  • D. 9.5 cm
Reveal answer and explanation for Question 1 (MAM-PP-01)

Answer: C. The CC's PNL should be within 1 cm of the MLO's PNL. 10.2 cm is 0.8 cm short, so it passes. D (1.5 cm short) and A (3.0 cm short) suggest posterior tissue was left off the CC. B is 1.5 cm longer, which also falls outside the 1 cm window, so that pair doesn't meet the criterion either.

Source: ACR Learning Network, Mammography Positioning Review Guidelines, Major criteria: Posterior Nipple Line. MAM-PP-01

Question 2 · MLO evaluation: pectoral muscle and nipple

Which MLO meets ACR positioning criteria?

  • A. The muscle reaches below the PNL, but the nipple isn't in profile on either view of that breast
  • B. The pectoral muscle is wider at the top, narrows inferiorly, and reaches the level of the PNL; the nipple is in profile on this view
  • C. The pectoral muscle is a thin, concave sliver that ends about 2 cm above the PNL
  • D. No pectoral muscle is visible, but the nipple is in profile
Reveal answer and explanation for Question 2 (MAM-PP-02)

Answer: B. A good MLO shows the pectoralis down to the PNL or below, wider superiorly and narrowing inferiorly. C describes inadequate muscle: narrow, concave, and more than 1 cm short of the PNL. D means posterior tissue is missing. A fails a separate major criterion: the nipple must be in profile on at least one view of each breast.

Source: ACR Learning Network, Mammography Positioning Review Guidelines, Major criteria: nipple in profile; minor criteria: amount of pectoralis major muscle. MAM-PP-02

Question 3 · Additional views: ML vs LM

A finding sits in the far medial breast. Which 90° lateral view puts it closest to the image receptor?

  • A. Mediolateral (ML)
  • B. Exaggerated craniocaudal lateral (XCCL)
  • C. Axillary tail
  • D. Lateromedial (LM)
Reveal answer and explanation for Question 3 (MAM-PP-03)

Answer: D. The LM places the medial breast against the image receptor. That keeps a medial lesion close to the receptor, which cuts object-to-image distance and the blur that comes with it. ML (A) is the choice for lateral lesions. XCCL (B) and the axillary tail view (C) are aimed at lateral and axillary tissue.

Source: IARC Atlas of Breast Cancer Early Detection, 90° lateral views. MAM-PP-03

Question 4 · Magnification

Which statement about magnification views is correct under MQSA equipment standards?

  • A. The system must be able to operate with the grid removed for magnification
  • B. A grid must stay in place to control scatter from the air gap
  • C. Any magnification factor is acceptable as long as it exceeds 1.1
  • D. Magnification capability is optional for systems used for problem-solving work
Reveal answer and explanation for Question 4 (MAM-PP-04)

Answer: A. Systems used for magnification must be able to work with the grid removed; the air gap handles scatter, so B is backward. Systems must offer at least one magnification value between 1.4 and 2.0, which rules out C. Systems used for noninterventional problem-solving procedures must have magnification available, so D is wrong.

Source: 21 CFR 900.12, (b)(4)(iii) and (b)(6). MAM-PP-04

Question 5 · Patient variance: breast implants

A screening patient has subpectoral breast implants, and there's no contraindication or physician order changing the exam. What does MQSA require?

  • A. Standard CC and MLO views with full compression on the implant
  • B. Referral for breast MRI instead of mammography
  • C. Views that maximize visualization of breast tissue, such as adding implant-displaced views to the standard views
  • D. Implant-displaced views only, because standard views are contraindicated with implants
Reveal answer and explanation for Question 5 (MAM-PP-05)

Answer: C. MQSA requires facilities to ask about implants before the exam and, unless contraindicated or modified by a physician, to use views that maximize visualization of breast tissue. The ARRT outline lists implant-displaced views and breast augmentation under patient variance. D and A each fail the regulation's requirement to maximize visualization of breast tissue. B substitutes another modality without support from the facts in the stem.

Source: 21 CFR 900.12, (g); ARRT Mammography Content Specifications, Procedures 2.A.10 and 2.B.9, p. 8. MAM-PP-05

Question 6 · Interventional procedures: infection control

After a stereotactic biopsy, there's blood on the compression paddle. What governs how the paddle is cleaned?

  • A. Any hospital-strength disinfectant, regardless of the equipment manufacturer's instructions
  • B. A dry gauze wipe now, with full disinfection at the end of the day
  • C. Whatever the radiologist performing the biopsy prefers
  • D. The facility's documented infection-control procedures, which must follow the manufacturer's cleaning and disinfection recommendations
Reveal answer and explanation for Question 6 (MAM-PP-06)

Answer: D. MQSA requires each facility to set up and follow a documented system for cleaning and disinfecting mammography equipment after contact with blood or other potentially infectious materials, in line with applicable regulations and the manufacturer's recommended procedures. A can damage equipment or break the manufacturer's instructions. B delays disinfection. C isn't a documented system.

Source: 21 CFR 900.12, (e)(13). MAM-PP-06

Question 7 · Localization: rolled CC views

A lesion is seen only on the CC view. For a rolled CC, the technologist rolls the superior half of the breast medially, and the lesion moves medially compared with the true CC. Where is the lesion?

  • A. In the axillary tail
  • B. In the superior half of the breast
  • C. In the inferior half of the breast
  • D. At the level of the nipple
Reveal answer and explanation for Question 7 (MAM-PP-07)

Answer: B. On a medially rolled CC, a lesion in the superior half moves medially with the roll, while a lesion in the inferior half appears more lateral. That's how rolled views localize a finding seen on only one standard view.

Source: UCLA Health Radiology, Breast Lesion Localization, Rolled CC views. MAM-PP-07

Question 8 · Imaging examinations: screening vs diagnostic

A patient arrives for imaging because she found a lump. Under MQSA definitions, what kind of exam is this?

  • A. Diagnostic: individualized views for a patient with symptoms or physical signs
  • B. Screening: routine views, because she has no prior abnormal mammogram
  • C. Screening, converted to diagnostic only if the radiologist finds something
  • D. Neither, because a palpable lump requires ultrasound instead
Reveal answer and explanation for Question 8 (MAM-PP-08)

Answer: A. MQSA defines a screening mammogram as routine views of a patient with no symptoms, and a diagnostic mammogram as individualized views of a patient with breast symptoms, physical signs of breast disease, or abnormal screening findings. A palpable lump is a symptom, so B and C are wrong. D skips the mammogram the referral asked for.

Source: 21 CFR 900.12, (f)(1). MAM-PP-08

Anatomy, physiology, and pathology (6 questions)

Question 9 · Localization: triangulation

A mass sits the same distance from the nipple on the MLO and on a 90° mediolateral (ML) view, but it appears higher on the ML than on the MLO. Where is it?

  • A. Lateral breast
  • B. Central breast
  • C. It can't be localized without ultrasound
  • D. Medial breast
Reveal answer and explanation for Question 9 (MAM-APP-01)

Answer: D. Going from the MLO to a true lateral, medial lesions move up and lateral lesions move down ("muffins rise, lead falls"). A central lesion barely changes height, so B is wrong. Triangulating with the CC, MLO, and a 90° lateral localizes it without ultrasound, so C is wrong.

Source: IARC Atlas of Breast Cancer Early Detection, 90° lateral views; UCLA Health Radiology, Breast Lesion Localization. MAM-APP-01

Question 10 · Reporting: BI-RADS assessment

A finding is assessed as BI-RADS category 3, "probably benign." What management does this category usually carry?

  • A. Additional imaging before an assessment can be given
  • B. Short-interval follow-up, typically a 6-month diagnostic mammogram
  • C. Tissue biopsy
  • D. Return to routine annual screening
Reveal answer and explanation for Question 10 (MAM-APP-02)

Answer: B. Category 3 findings carry less than a 2% likelihood of malignancy and are usually managed with short-interval follow-up, often at 6 months. Biopsy (C) is for categories 4 and 5. Routine screening (D) fits categories 1 and 2. A describes category 0, "incomplete."

Source: ACR BI-RADS current release; Breast Imaging Reporting and Data System v2025: Key Updates in Mammography, Category 3. MAM-APP-02

Further citation details: Current release: BI-RADS v2025; Category 3 (Probably Benign)

Question 11 · Calcifications: milk of calcium

Calcifications look like smudgy, round densities on the CC view. Which view best shows whether they're milk of calcium?

  • A. A 90° lateral (ML or LM) view
  • B. A spot compression CC
  • C. A repeat MLO with more compression
  • D. A cleavage view
Reveal answer and explanation for Question 11 (MAM-APP-03)

Answer: A. Milk of calcium settles in the dependent part of tiny cysts. A horizontal-beam 90° lateral shows it as sharp, curved or linear "teacup" shapes, while the CC sees it from above as smudges. The other views don't use a horizontal beam, so they can't show the layering. The ARRT outline lists milk of calcium among typically benign calcifications.

Source: IARC Atlas of Breast Cancer Early Detection, 90° lateral views; ARRT Mammography Content Specifications, Procedures 1.E.1.d, p. 7. MAM-APP-03

Question 12 · Pathology: benign with upgrade potential

Which lesion does the current ARRT outline group as "benign with upgrade potential" rather than malignant?

  • A. Invasive lobular carcinoma
  • B. Paget disease of the breast
  • C. Lobular carcinoma in situ (LCIS)
  • D. Ductal carcinoma in situ (DCIS)
Reveal answer and explanation for Question 12 (MAM-APP-04)

Answer: C. Despite the name, LCIS is grouped with ADH, ALH, papilloma with atypia, flat epithelial atypia, radial scar, and phyllodes tumor as benign with upgrade potential. DCIS, invasive lobular carcinoma, and Paget disease are listed as malignant.

Source: ARRT Mammography Content Specifications, Procedures 1.E.3 and 1.E.4, p. 7. MAM-APP-04

Question 13 · Localization: clock face and quadrants

A finding is reported at 10 o'clock in the right breast. Which quadrant is it in?

  • A. Lower inner
  • B. Upper outer
  • C. Upper inner
  • D. Lower outer
Reveal answer and explanation for Question 13 (MAM-APP-05)

Answer: B. Clock positions are read as if you're facing the patient, so the two breasts mirror each other. In the right breast, the outer half runs from 6 to 12 o'clock and the upper half from 9 to 3, so 10:00 is upper outer. In the left breast, 10:00 would be upper inner (C).

Source: UCLA Health Radiology, Breast Lesion Localization, Quadrants on CC and MLO. MAM-APP-05

Question 14 · Calcifications: suspicious morphology

Which calcification descriptor does the ARRT outline list under suspicious morphology?

  • A. Milk of calcium
  • B. Vascular
  • C. Coarse
  • D. Fine pleomorphic
Reveal answer and explanation for Question 14 (MAM-APP-06)

Answer: D. The outline lists amorphous, heterogeneous, and fine pleomorphic as suspicious morphology. Skin, vascular, coarse, milk of calcium, and dystrophic calcifications are listed as typically benign.

Source: ARRT Mammography Content Specifications, Procedures 1.E.1.d, p. 7. MAM-APP-06

Image production (7 questions)

Question 15 · Mammographer QC: phantom image

A facility follows the 2018 ACR Digital Mammography QC Manual. How often does the QC technologist acquire and score the ACR digital mammography phantom on each unit?

  • A. Weekly
  • B. Monthly
  • C. Quarterly
  • D. Semiannually
Reveal answer and explanation for Question 15 (MAM-IP-01)

Answer: A. ACR's unit checklist for the 2018 manual lists phantom image quality as weekly. Monthly (B) is the compression thickness indicator, visual checklist, and acquisition workstation monitor QC. Quarterly (C) is the facility QC review. Semiannual (D) is compression. The phantom passes with at least 2 fibers, 3 speck groups, and 2 masses.

Source: ACR Digital Mammography Unit QC Summary Checklist; ACR webinar: The 2018 ACR Digital Mammography QC Manual (Berns), Phantom pass criteria slide. MAM-IP-01

Question 16 · Mammographer QC: corrective action

On Monday morning, the weekly phantom image fails to reach the minimum score. What should happen next?

  • A. Repeat the phantom at next week's scheduled test
  • B. Continue diagnostic exams only until service arrives
  • C. Find and fix the problem before performing any more patient exams on that unit
  • D. Keep screening and finish corrective action within 30 days
Reveal answer and explanation for Question 16 (MAM-IP-02)

Answer: C. Under MQSA, a failed phantom image quality test, and failures in a digital unit's QC program, must be corrected before more exams are performed with that component. The 30-day window in D applies only to tests that aren't on that list. A and B keep imaging patients on equipment that just failed an image-quality test.

Source: 21 CFR 900.12, (e)(8)(ii)(A)-(B). MAM-IP-02

Question 17 · Equipment: compression force

MQSA's compression-device standard limits the maximum force of the initial power-driven compression. During the semiannual test, a unit's power drive reaches 48 pounds. Which statement is correct?

  • A. It passes; the only requirement is at least 25 pounds of force
  • B. It passes, because manual fine adjustment can always exceed the power drive
  • C. It fails; the initial power drive must not exceed 35 pounds
  • D. It fails; the initial power drive must max out between 25 and 45 pounds (111–200 newtons)
Reveal answer and explanation for Question 17 (MAM-IP-03)

Answer: D. The regulation requires compression of at least 111 N (25 lb) and an initial power-drive maximum between 111 N (25 lb) and 200 N (45 lb). A states only half the rule. B confuses fine adjustment with the power-drive limit. C uses a limit that doesn't exist.

Source: 21 CFR 900.12, (e)(4)(iii); ACR Digital Mammography Unit QC Summary Checklist, Compression (semiannual). MAM-IP-03

Question 18 · Mammographer QC: frequencies

In the 2018 ACR Digital Mammography QC program, which technologist test is performed quarterly?

  • A. Acquisition workstation monitor QC
  • B. Facility QC review
  • C. Compression thickness indicator
  • D. Compression force
Reveal answer and explanation for Question 18 (MAM-IP-04)

Answer: B. The facility QC review is the quarterly item. The compression thickness indicator (C) and acquisition workstation monitor QC (A) are monthly; compression (D) is semiannual. Repeat analysis is listed as optional and as needed in the 2018 digital program, which trips up anyone who learned the old film-screen schedule.

Source: ACR Digital Mammography Unit QC Summary Checklist. MAM-IP-04

Question 19 · Medical physicist tests: average glandular dose

Under MQSA, the average glandular dose from a single craniocaudal view of an FDA-accepted phantom simulating a standard breast may not exceed:

  • A. 5.0 mGy (0.5 rad)
  • B. 10 mGy (1 rad)
  • C. 3.0 mGy (0.3 rad)
  • D. 1.0 mGy (0.1 rad)
Reveal answer and explanation for Question 19 (MAM-IP-05)

Answer: C. The limit is 3.0 mGy per exposure, measured by the medical physicist using the facility's clinical technique for a standard breast. The same maximum applies to digital and other non-screen-film systems. The outline expects only basic knowledge of physicist tests, and dose is one worth memorizing.

Source: 21 CFR 900.12, (e)(5)(vi) and (e)(6). MAM-IP-05

Question 20 · MQSA: breast density reporting

A mammography report classifies breast density as "There are scattered areas of fibroglandular density." What must the patient's lay summary say about density?

  • A. That her breast tissue is not dense, with the standard explanation and advice to talk to her provider
  • B. That her breast tissue is dense and that other imaging tests may help find cancers
  • C. Nothing; density only has to be reported when the breasts are dense
  • D. That she should schedule a supplemental ultrasound
Reveal answer and explanation for Question 20 (MAM-IP-06)

Answer: A. Every lay summary now includes a density statement. "Almost entirely fatty" and "scattered areas of fibroglandular density" trigger the not-dense statement. "Heterogeneously dense" and "extremely dense" trigger the dense statement, which adds that other imaging tests may help some people (B). C is wrong under the current federal rule because every lay summary includes a density statement. D is a clinical decision, not required lay-summary wording.

Source: 21 CFR 900.12, (c)(1)(vi) and (c)(2)(iii)-(iv). MAM-IP-06

Question 21 · MQSA: recordkeeping

A patient has a mammogram and never returns to the facility. Unless state law requires longer, how long must the facility keep her original mammograms and reports?

  • A. 24 months
  • B. At least 10 years
  • C. At least 5 years
  • D. At least 3 years
Reveal answer and explanation for Question 21 (MAM-IP-07)

Answer: B. Original mammograms and reports must be kept for the longest of: at least 5 years; at least 10 years if the facility performs no more mammograms on the patient; or any longer state or local requirement. C is the minimum when the patient keeps coming back. D is how long serious consumer complaints are kept, and A is how long records of departed staff are kept. Both are real MQSA numbers, just for different records.

Source: 21 CFR 900.12, (c)(4)(i), (h)(2), (a)(4). MAM-IP-07

Patient care (4 questions)

Question 22 · Patient education: screening guidelines

A 47-year-old patient has no personal or strong family history of breast cancer, no known gene mutation, and no chest radiation before age 30. She asks how often the American Cancer Society recommends screening mammograms at her age. What's the best answer?

  • A. Every other year
  • B. Screening isn't recommended until age 50
  • C. Every year, along with a yearly clinical breast exam
  • D. Every year
Reveal answer and explanation for Question 22 (MAM-PC-01)

Answer: D. ACS recommends yearly mammograms for average-risk women ages 45 to 54. Every other year (A) is an option ACS gives at 55 and older. B isn't ACS guidance; ACS even gives women 40 to 44 the option to start yearly screening. C adds something ACS doesn't recommend: clinical breast exams aren't recommended for screening average-risk women at any age.

Source: American Cancer Society screening recommendations, Average-risk guidelines; CBE and BSE. MAM-PC-01

Question 23 · Patient communication: results

During a screening mammogram, the patient asks, "Do you see anything bad?" What's the most appropriate response?

  • A. Tell her you aren't allowed to talk about her exam, and move on
  • B. Point out the dense areas on the monitor and explain what they might mean
  • C. Explain that a radiologist will interpret the images and tell her how and when she'll get her results
  • D. Reassure her that the images look normal so far
Reveal answer and explanation for Question 23 (MAM-PC-02)

Answer: C. The ARRT outline separates the mammographer's job, explaining how results arrive and why more imaging might be needed, from the clinician's role in explaining results and findings. D and B interpret images, which isn't the technologist's role, and a "looks normal" comment can turn out wrong. A is right about scope but skips the patient education. Federal rules back up the process: the facility must send a lay summary within 30 days, or within 7 days if the assessment is Suspicious or Highly Suggestive of Malignancy.

Source: ARRT Mammography Content Specifications, Patient Care A.4, p. 2; 21 CFR 900.12, (c)(2). MAM-PC-02

Question 24 · Patient assessment: risk factors

The current ARRT outline sorts breast cancer risk factors into inherent and social categories. Which of these is listed as an inherent risk factor?

  • A. Previous chest radiation
  • B. Tobacco use
  • C. Excessive alcohol use
  • D. Body weight
Reveal answer and explanation for Question 24 (MAM-PC-03)

Answer: A. The outline lists previous chest radiation under inherent risk factors, along with age, personal and genetic history, and breast density. B, C, and D fall under social (lifestyle) risk factors. ACS also counts chest radiation before age 30 as a high-risk factor.

Source: ARRT Mammography Content Specifications, Patient Care B.1, p. 2; American Cancer Society screening recommendations, High-risk guidelines. MAM-PC-03

Question 25 · Patient history: implant inquiry

Federal MQSA regulations require every facility to have a procedure for asking each patient about which of the following before the mammogram?

  • A. Caffeine intake
  • B. Breast implants
  • C. Deodorant or powder use
  • D. Date of last menstrual period
Reveal answer and explanation for Question 25 (MAM-PC-04)

Answer: B. MQSA requires a procedure to ask whether the patient has breast implants before the exam. Removing deodorant (C) is a standard pre-exam instruction and appears in the ARRT outline, but it isn't the inquiry MQSA mandates. D and A aren't MQSA requirements.

Source: 21 CFR 900.12, (g)(1). MAM-PC-04

How to use your result

Count your misses by section, not just your total. Two or more misses in one section is your cue to review that topic before doing more random questions. That's our study rule of thumb, not an ARRT rule, and with only 4 to 8 items per section it points you to a topic rather than measuring your ability.

Keep a short error log as you go:

Question Section Why I missed it The rule in one sentence Source I checked Retest date
Question 1 Positioning Forgot the 1 cm PNL rule CC PNL must be within 1 cm of MLO PNL ACR positioning guidelines Day 6

"Why I missed it" usually falls into one of five buckets: didn't know the fact, misread the stem, mixed up geometry (which way a lesion moves), confused two rules (MQSA vs ACR, film-screen vs digital), or rushed.

Your study plan

Pick the timeline that matches your exam date. Both follow the same priority order as the table at the top. They're editorial plans built from ARRT's blueprint, not an ARRT requirement or a guaranteed outcome.

Two weeks, 60 to 90 minutes a day

Each day: about 45 to 60 minutes learning the topic, then 15 to 30 minutes of retrieval (close your notes and write or say what you remember, then check).

Day Focus Done when you can…
1 Baseline Answer all 25 questions untimed, start your error log, and read the ARRT outline once end to end
2 Patient care Explain ACS screening guidance by age, inherent vs social risk factors, how results reach the patient, and basic treatment terms
3 Anatomy and localization Name breast anatomy down to the TDLU, and convert clock positions to quadrants for either breast
4 Pathology I Describe mass shape/margin/density and sort calcifications into typically benign vs suspicious morphology
5 Pathology II and reporting Match BI-RADS categories 0–6 to their management, list the four density categories, and sort lesions into benign, upgrade-potential, and malignant
6 Positioning I: CC and MLO State the CC and MLO criteria (PNL rule, pectoralis, nipple in profile, IMF) and the fix for each common miss
7 Positioning II: added views Say what each outline view shows and when you'd choose it: ML, LM, XCCL/XCCM, cleavage, axillary tail, tangential, rolled
8 Positioning III: problem patients Handle implants, magnification and spot compression, kyphosis, pectus, devices, male patients, very large or small breasts
9 Digital imaging and informatics Explain FFDM vs DBT vs synthesized images, acquisition vs radiologist workstations, PACS storage and troubleshooting, CAD
10 Technique and image evaluation Connect kVp, mAs, AEC, target/filter, focal spot, grid, and compression to image quality; name EQUIP's three areas
11 Quality control Give purpose, frequency, and failure action for every mammographer QC test; give purpose and frequency for physicist tests
12 MQSA Recite the MQSA numbers in the table below, the assessment categories, and the density lay-summary rule
13 Interventional procedures + weak spots Walk through biopsy prep (history, consent, time-out), stereotactic vs ultrasound biopsy, localization and specimen imaging; redo your error-log items
14 Timed review Redo all 25 at about one minute each, review every rationale, confirm your Pearson VUE appointment and two IDs

Six weeks, about 5 to 6 hours a week

Week Focus Do this
1 Baseline + patient care Days 1–2 above, then a second pass on anything you missed
2 Anatomy and localization Day 3 above; drill triangulation and rolled views until they're automatic
3 Pathology and reporting Days 4–5 above
4 Positioning Days 6–8 above; this is the biggest single category, so give it the most hours
5 Image production Days 9–12 above
6 Procedures + timed review Days 13–14 above, plus a full redo of your error log

Only one week? Do Days 1, 6–8, 11–12, and 14 in that order. That covers positioning, QC, and MQSA first, then timed review.

What the outline actually expects

ARRT's content specifications include four footnotes that change how deep you need to go:

  • Positioning: You're expected to know positioning as presented in the ACR's 1999 Mammography Quality Control Manual, Clinical Image Quality section. ARRT notes that about six items in the positioning category cover the standard CC and MLO. The rest of the 39 go to added views, patient variance, imaging exams, and procedures. (p. 8, footnote 3)
  • Breast ultrasound, breast MRI, and interventional procedures: basic knowledge. (p. 8, footnote 4)
  • Quality control: detailed understanding of every mammographer QC test (purpose, frequency, equipment and procedure, performance criteria, documentation, corrective action) and basic understanding of medical physicist tests (purpose and frequency), as referenced in the 2018 ACR Digital Mammography QC Manual. (p. 5, footnote 2)
  • Treatment options: definitions and basic descriptions only. (p. 3, footnote 1)

What changed on September 1, 2025

If your review book or deck predates September 2025, patch it against this list from ARRT's announcement:

  • Added: patient education about how results are delivered; more detail on inherent and social risk factors; synthesized imaging; PACS troubleshooting; breast density reporting requirements under MQSA; EQUIP and patient-related artifacts; more detail on patient preparation for interventional procedures.
  • Removed: TRAM and latissimus dorsi flaps; sentinel node mapping; cyst aspiration; fine-needle aspiration biopsy.
  • Unchanged total: still 115 scored questions; unscored pilots rose from 25 to 30.

High-yield quick tables

Mammographer QC under the 2018 ACR Digital Mammography QC Manual

Test Frequency
ACR digital mammography phantom image quality (pass: at least 2 fibers, 3 speck groups, 2 masses) Weekly
CR cassette erasure (only if CR is used) Weekly
Compression thickness indicator Monthly
Visual checklist Monthly
Acquisition workstation monitor QC Monthly
Facility QC review Quarterly
Compression Semiannual
Manufacturer detector calibration Per the manufacturer
Repeat analysis Optional, as needed

Frequencies from the ACR Digital Mammography Unit QC Summary Checklist; phantom pass criteria from ACR's 2018 QC manual webinar for technologists. Radiologist workstations and other display devices are tracked on ACR's separate display-device checklist.

Two traps. A facility can follow its equipment manufacturer's QC program instead of the ACR manual, and those frequencies can differ, so read the stem for which program applies. And many older review sources still teach film-screen schedules like daily processor QC and quarterly repeat analysis. Those aren't the 2018 digital ACR schedule.

MQSA numbers worth memorizing

All from 21 CFR 900.12, the federal MQSA quality standards.

Rule Number Section
Technologist initial mammography training 40 contact hours, including 25 exams under direct supervision; 8 hours in each modality used (a)(2)(ii)
Technologist continuing education 15 CEUs in mammography per 36 months, at least 6 per modality used (a)(2)(iii)
Technologist continuing experience 200 mammography exams per 24 months (a)(2)(iv)
Before using a new modality independently 8 hours of training (a)(2)(iii)(E)
Lay summary to the patient Within 30 days; within 7 days if Suspicious or Highly Suggestive of Malignancy (c)(2)
Follow-up report after "Incomplete: need prior mammograms" Within 30 days of the first report (c)(1)(v)(B)
Keep original mammograms and reports At least 5 years; at least 10 years if no more mammograms at the facility; longer if state law requires (c)(4)(i)
Transfer originals on request Within 15 days (c)(4)(ii)
Average glandular dose, standard-breast CC No more than 3.0 mGy (e)(5)(vi), (e)(6)
Compression force At least 25 lb; initial power drive maxes out between 25 and 45 lb (111–200 N) (e)(4)(iii)
Magnification Grid removable; at least one factor between 1.4 and 2.0 (b)(4)(iii), (b)(6)
Medical outcomes audit At least every 12 months (f)(2)
Serious consumer complaints kept 3 years (h)(2)

MQSA's report assessment categories are word phrases: Negative, Benign, Probably Benign, Suspicious, Highly Suggestive of Malignancy, Known Biopsy-Proven Malignancy, and Post-Procedure Mammogram for Marker Placement, plus two "Incomplete" statements. Its four density statements run from "almost entirely fatty" to "extremely dense, which lowers the sensitivity of mammography." The density lay-summary wording became federally required with amendments effective September 10, 2024.

Positioning and localization rules

Rule What to remember Source
Nipple in profile On at least one view of each breast ACR Learning Network
PNL check CC PNL within 1 cm of MLO PNL ACR Learning Network
Adequate MLO pectoralis Reaches the PNL or below; wider at the top, narrowing down; not concave ACR Learning Network
Triangulation (MLO → 90° lateral) Medial lesions rise, lateral lesions fall, central lesions stay put IARC Atlas
Which 90° lateral ML for lateral lesions; LM for medial lesions IARC Atlas
Milk of calcium Layers into "teacups" on a 90° lateral; smudgy on CC IARC Atlas
Rolled CC Superior lesions move with the roll; inferior lesions move the opposite way UCLA Radiology
Clock face Read facing the patient: right 10:00 = upper outer; left 10:00 = upper inner UCLA Radiology

Exam-day facts that change how you practice

Fact Detail
Questions 115 scored + 30 unscored pilots (145 in all); pilots aren't marked
Time 150 minutes of test time in a 170-minute appointment (8-minute tutorial, 2-minute nondisclosure agreement, 10-minute survey)
Pace About one minute per question
Navigation You must answer each question before moving on. If unsure, guess, flag it, and return if time allows. Blank answers can't help you.
Order Questions appear in random order, not grouped by category
Passing Scaled score of 75 on a 1–99 scale. That's not 75% correct; ARRT sets the raw-score equivalent for each exam form.
Section scores Reported from 0.1 to 9.9; useful for spotting weak areas, but less reliable than the total

Sources: 2026 Postprimary Handbook, pp. 39, 45, 47, 56–57; ARRT: How to Interpret Your Score Report. For what happens after test day, see our guide to ARRT exam results.

Eligibility in brief

ARRT's Mammography credential is earned only through the postprimary pathway. You need:

  • A supporting credential. ARRT lists Radiography as the supporting discipline for Mammography (handbook p. 6).
  • Structured education. At least 16 credits covering the content outline, with at least one in each major category, all earned no more than 24 months before you apply (handbook pp. 8–9).
  • Clinical experience. Independently complete mammograms on 75 patients; participate in a set number of each QC test; review at least 10 cases with an MQSA-qualified interpreting physician; and observe, assist with, or take part in at least five elective procedures. Procedures count only after you're MQSA compliant (Mammography Clinical Experience Requirements).
  • Application. $225 for the postprimary pathway (ARRT fees). Your exam window runs 365 calendar days, and you get three attempts within three years of your first window opening (handbook pp. 18, 32).

An ARRT credential and MQSA qualification are separate things, and neither one is a state license. Check your state's requirements for permission to practice.

Common questions

Does ARRT recommend a review book or course? No. ARRT doesn't endorse review programs, mock registries, or study guides, and warns against materials claiming to contain real ARRT questions. It points candidates to the content specifications (handbook p. 40).

What if I don't pass? You can reapply, within the three-attempts-in-three-years limit. Before a retake, compare the current content specifications with the ones you studied (handbook p. 60). Details: ARRT exam retake policy.

What happens after I pass? You renew every year, complete continuing education every two years, and meet Continuing Qualifications Requirements every ten years (handbook p. 61). See ARRT continuing education requirements.

Ready to schedule? Once your Candidate Status Report posts, wait 24 hours, then book with Pearson VUE. Here's how ARRT exam registration works.

Sources

Last verified October 27, 2026. We checked ARRT's Mammography content specifications, 2026 postprimary handbook, score-report guidance, application fees, and Mammography clinical experience requirements; 21 CFR 900.12; ACR QC forms and positioning guidance; and the teaching sources cited in each rationale.

Source access note: the IARC Atlas of Breast Cancer Early Detection link cited in Questions 3, 9, and 11 (and in the positioning rules table) could not be reached during our latest link check. We've kept the original citation unchanged.

About this page. Castleport Test Prep is an independent exam prep publisher. We aren't affiliated with, endorsed by, or approved by The American Registry of Radiologic Technologists (ARRT). Exam and credential names identify their subjects, and trademarks belong to their respective owners. The practice questions here are original and aren't ARRT exam questions. Using this page doesn't guarantee a passing score, certification, or employment. By the Castleport Test Prep Editorial Team.